Chapter 7. Challenges in children; critical choices

7.1 Introduction; children should not be treated as young adults

Counselling adults, who should profit from amplification according to their audiogram, is not always successful. Adults tend to deny their hearing problems. If their attitude is more positive, they may only be interested in revalidation with barely visible devices, even if speech recognition with such devices is not optimal.

For children, there is less room to move, because the better the hearing the greater the chance that the child will develop normally. According to the ‘Practice Guidelines’ (for the placement of percutaneous devices in children with conductive hearing loss; Roman et al., 2011), the aided thresholds should be 17.5 dB HL or better, which is close to the limit of 15 dB HL proposed by Northern & Downs (1991, their chapter 1). The latter authors argued that young children need (sub)normal hearing to ensure normal speech and language development. Therefore, when counselling the parents of a child with conductive hearing loss, sufficiently powerful amplification options should be recommended. The main purpose of (any) hearing aid fitting is to optimize the audibility of speech; a popular tool to assess the audibility of speech, based on the audiogram, is the Speech Intelligibility Index (SII; e.g., Amlani et al., 2002).

Intermezzo. The Speech Intelligibility Index.

The SII is the proportion of normal speech that is audible to a hearing-impaired person fitted with a hearing device; the SII ranges from 0 to 100, 0 means that normal speech is not audible and 100 means that the individual can hear normal speech fully. The SII is based on the spectrum of conversational speech with a +/- 15 dB range, often referred to as the speech banana. The speech banana can be represented in the audiogram, see Figure 7.1, the grey area. To calculate the SII, for each (1/3 octave) frequency band, the proportion of speech that is audible is weighted according to the contribution of that frequency band to speech intelligibility (weighting factors are indicated by the density of the dots in the figure). To obtain the SII, the dots above the hearing thresholds must be counted for all 1/3 frequency bands (Mueller and Killion, 1990; Killion and Mueller, 2010). In the figure, post-intervention hearing thresholds are presented from two different treatments. Counting the dots shows that the SII is 90 and 22, respectively, for intervention 1 and 2. Killion & Mueller (2010) also published a graph with the relationship between the SII and word recognition scores (WRS; their Figure 3). According to that graph, the SII of 90 and 22 corresponds to word recognition scores of 95% and 20%, respectively. End Intermezzo

With reference to Figure 7.1, treatment 1 is clearly much more effective than treatment 2. These two ‘intervention’ curves come from a study by Farnoosh et al. (2014). They compared post-intervention hearing thresholds of two groups of children with congenital aural atresia; group 1 (intervention 1) consists of children who use a pBCD, and group 2 (intervention 2), consists of children with surgical ‘atresia repair’. The pre-intervention audiograms of the two groups were similar. The figure shows that the ‘maximum 17.5 dB hearing loss criterion’ (Practice Guidelines’; Roman et al., 2011) is almost met. It should be noted here that bilateral treatment should always be considered because in addition to better hearing in noise and spatial hearing, binaural summation leads to better hearing thresholds, in the order of 3 to 5 dB HL (see Chapter 6.1). As an example, Liu et al. (2023) studied children with bilateral conductive hearing loss who were fitted with two passive tBCDs and showed that indeed the aided PTA4 improved by 4 dB and the word recognition by 9% when both instead of one were used.

Importantly, statistics showed that young children with bilateral hearing devices developed better cognitive abilities compared to those using a unilateral hearing aid, which was attributed to better perception of soft (speech) sounds and better speech perception in noisy places (de Raeve et al., 2015). The factor between better hearing and better cognitive development is incidental learning. Most of the things children learn are learned in informal everyday situations. Only about a third of their knowledge is acquired in structured, quiet settings (e.g. at school). The remaining 2/3 is acquired during the day while playing, watching TV, talking with parents and other persons, and so on (de Raeve et al., 2015). Especially under these circumstances, the hearing-impaired child benefits from bilateral hearing devices. The observation that children with unilateral hearing loss (second ear normal) may have delayed educational development, probably has the same cause (Kuppler et al., 2013). This reasoning emphasizes that hearing with two ears is of utmost importance for the development of speech, language and cognition in children.

7 punt 1

Figure 7.1. Audiogram format with speech area according to Mueller and Killion (gray area). Post-intervention thresholds are shown of two different treatments, 1 (red line) and 2 (green line). The SII of the two interventions is determined by counting the dots above the hearing threshold (Source audiogram: Mueller and Killion, 1990).

7.2 Literature review, the effectiveness of interventions for children with conductive hearing loss

Referring to Section 3.2, in patients with predominant conductive hearing loss, the best aided thresholds were obtained with pBCDs while the worst were found with passive tBCDs. The other devices were in between. A similar review of the literature was carried out, with additionally the search term ‘children’. Papers were included if 1. two interventions were compared, 2. at least 10 children were involved, 3. with a unilaterally fitted device. Search period was 2008-2020. See note at the end of this chapter for references. Virtually all studies involved children with congenital aural atresia.

Table 7.1 shows the mean PTA4bc and the mean post-intervention PTA4 of the included papers; the mean PTA4bc are quite similar. The mean post-intervention PTA4 ranged from 19 dB HL and 41 dB HL, with, again, the best result for the pBCD: the 19 dB HL, close to the target value of 17.5 dB HL (‘Practise Guidelines’, Roman et al., 2011) and the 15 dB HL introduced by Northern & Downs (1991). The Table also presents the effective gain (EG) which is negative for all treatments, indicating a ‘remaining air-bone gap’, despite the interventions.

Using the studies that presented frequency-specific thresholds, the mean threshold by frequency and type of intervention, was calculated to determine the SII. The estimated WRS (in multiples of 5%) based on the SII, is presented in the Table. Once again, also these WRS values show that the interventions are not equivalents at all. The result after surgical correction of atresia is particularly disappointing, as is that after application of passive tBCDs. Moreover, the variability in the mean post-intervention PTA4 values is quite large, ranging from 7 dB to 16 dB, which may be clinic related.

Table 7.1

Intervention

 

N

studies

N

children

Mean PTA4bc

dB HL

Mean post-intervention PTA4 with (range)

dBHL

Estimated WRS; % correct Effective gain

dB

Atresia repair 5 199 11 41 (33-47) 30% -30
Passive tBCD 10 168 13 34 (27-43) 45 -21
‘Bonebridge’ 7 100 10 27 (19-31) 85 -17
pBCD 6 135 11 19 (15-22) 95 -8
VSB 4 60 11 31 (27-40) 80 -20

 

7.3 Age at intervention and stability

During recent years, few studies are published that promote reconstructive surgery (atresia repair). Only relatively mild atresia cases should be considered for surgery (e.g., Declau et al., 1999). Atresia repair is becoming obsolete because of the development of (more) effective solutions: BCDs and MEIs.

For young children, babies, toddlers, few amplification options are available, after a diagnosed bilateral conductive hearing loss. When the tympanic membrane is visible, BTEs can be used. A passive tBCD, for example with softband, can be used in all cases. The BCD with softband is easier to use and thus better accepted than a steel headband (Hol et al., 2005). The position of the stimulating bone-vibrator over the skull can be easily changed. The BCD-aided thresholds with such devices are typically between 25 dB HL and 35 dB HL, which may enable learning speech at a basic level (Northern & Downs, 1991; Verhagen et al., 2008). Therefore, the replacement of a passive tBCD with a pBCD or another, more powerful solution must remain on the agenda while the child’s speech and language development must be monitored in the meantime.

The use of a pBCD implies implantation of a skin-penetrating titanium implant, anchored in the skull, behind the ear. Regarding the youngest age at which such implantation is possible, skull thickness plays an important role. At least 3 mm is preferred (Snik et al., 2005). This implies that the child must be approximately 4 years old; yet, even then, the loss of these implants is about three times higher than in adults (e.g., Dun et al., 2010; McDermott et al. 2009). Kruyt et al. (2020) published a literature review on medical complications in children with pBCDs. Implant loss occurred in 13,3% and revision surgery was needed in 16,8% of the implants. However, dividing the reviewed studies into those with the original implants/surgical procedures and those with upgraded implants/procedures showed an ‘improvement’ in implant loss from 17,1% to 5,9%. Of course, implant loss in children is a problem, but major steps forward have been made (see also Chapter 5). The pBCDs have been approved by the FDA for children aged 5 years and older.

Another option may be a passive tBCD with magnetic coupling. The subcutaneous magnet can be placed from age 5 onwards, according to the manufacturers. Souza et al. (2022) conducted a literature review on complications with passive tBCDs with magnetic coupling (in children and adults; a total 657 users). They concluded that complications, mostly skin problems, were common (25% complication rate), of which 2,6% were major complications leading to non-use. This questions the use of such devices, moreover, since these devices are no more powerful than passive tBCDs with softband (see for example den Besten et al., 2019).

VSB with its transducer (FMT) placed in the ‘round-window’ niche of the cochlea has been applied in children with congenital aural atresia, as little as 2 months (Mandala et al., 2011), but an audiological evaluation was lacking. Leinung et al. (2017) presented results obtained in 13 young children with, remarkably, unilateral atresia of the ear canal. The age at implantation ranged from 1.3 to 4.2 years with an average of 2.5 years. Four children became non-users. The mean PTA4bc of the remaining nine device users was 8 dB HL, the mean aided PTA4 was 40 dB HL, which is quite poor compared to the data presented in Table 7.1 and, of course, worse than the other, normal hearing ear. More evidence is needed to support VSB implantation in young children. However, data of seven studies in children with congenital aural atresia, implanted at age > 5 years, are presented in the Table 7.1. An average aided PTA4 of 31 dB HL was found.

On the application of active tBCDs in children, only data on Bonebridge implantation were found. The implanted part of the Bonebridge was 7.8 mm thick, therefore, a relatively deep bone bed had to be made. This is problematic in children because of thin skulls. Ratusniak et al (2019) concluded that the seize of the implant was problematic for use in children. Rather recently, another design was developed in which the implanted part was only partially embedded in the skull bone. The Bonebridge is approved by the FDA for individuals of 12 years and older while the VSB is approved for adults only. Nevertheless, the Bonebridge has been successfully applied in children aged 5 years and over; outcomes from 7 articles are presented in Table 7.1.

As for the VSB and Bonebridge, no prospective longitudinal studies targeting medical complications in children were found. However, a longitudinal study in adults has been published (Vickers et al., 2023); see section 5.1. Implant loss over time was not better than what was reported for pBCDs, for both devices, VSB and Bonebridge.

Atresia repair is becoming obsolete and is being replaced by hearing device fittings. Based on the capacity of the hearing devices, the pBCD is the best option. On major medical adverse events, no convincing evidence was found that one specific device was significantly worse than the others. For optimal development of the child, it is recommended to fit hearing devices bilaterally, irrespective of the device used.

Note. Studies included in the reviews leading to Table 7.1: Claros & Pujol, 2013; Frenzel et al., 2015; Verhagen et al., 2008; Farnoosh et al., 2014; Ricci et al., 2011; Denoyelle et al., 2015; Bouhabel et al., 2012; Evans & Kazahaya, 2007; Christensen et al., 2010; Mandala et al., 2011; Zhao et al., 2016; Baumgartner et al., 2016; Fan et al., 2019, Bravo-Torres et al., 2018, Kulasegarah et al., 2018, Ren et al., 2019, Ahn et al., 2018, Zernotti et al., 2019, Ratuszniak et al., 2019.

Chapter 5. (Post) surgical issues and cosmesis

5.1 Stability (implant survival rates) and skin problems

Systematic longitudinal studies on implant stability are relatively scarce but not for pBCDs. Reported disadvantages of percutaneous implants are implant loss and skin irritation around the implant, which may require treatment. It should be noted that fewer complications have occurred over the years thanks to more than 35 years of experience with these implants. New implantation procedures have been introduced as well as optimized percutaneous implants. Lagerkvist et al. (2020) reviewed the literature, search period 2009-2019. Twenty-seven pBCD studies were selected. Using their data, the implant survival rate at 1-1.5 years post-surgery was 98% (eight studies, 372 patients) and at 3-5 years post-surgery: 98,5% (3 studies, 130 patients). In other words, with the more recent percutaneous implants and surgical techniques, the survival rate of percutaneous implants is high, around 98%. Interestingly, more recently, Teunissen et al. (2024) published a longitudinal study on implant loss; such studies are important because implant loss may depend on follow-up. They included 701 consecutive patients with 807 implants, children and adults, also those with intellectual disabilities. This patient group is probably the most representative group of BCD users ever studied. Patients were implanted between 2009 and 2020 using the conventional linear incision surgical procedure or MIPS (see below). The survival rates at 1, 2 and 3 years of follow-up were 97,8%, 97% and 96%, respectively, which is somewhat worse than the data reported by Lagerkvist et al. This data will be used as a reference.

Longitudinal studies on implant survival of an active tBCD, the Bonebridge, have been published by Brkic et al. (2019) and Vickers et al. (2023). Brkic et al. presented data of 46 patients: the implant survival rate was 95,6 % at 1 year follow-up and 88% at 3 years. According to the authors, learning curves might have played a role. The study by Vickers et al. is rather unique in that it followed the strict NHS recommendations. The design of the clinical trial was published in 2018 (Vickers et al., 2018). The actual study was a non-sponsored, multi-centre longitudinal, observational study. The outcomes of the (in 2018) announced study were published in 2023 (Vickers et al., 2023). The reported implant survival rates (109 Bonebridge users), at 1year post-surgery was 97% and at 3 years, 93%. Obviously, these data are better than those reported by Brkic et al.

Longitudinal studies on VSB implant survival have been published by Sprinzl et al. (2021) and, again, Vickers et al (2023). Sprinzl et al. presented data of 46 adult patients, all with chronic ear infections (with or without cholesteatoma). The implant survival rate was 90% at 1year post-surgery and 73% at 4 years post-surgery. For the patients with chronic infections and cholesteatoma, the survival rate 4 years post-surgery was 60% while for the other patients it was 100% (without cholesteatoma). Vickers et al. also studied the VSB (163 VSB users); they reported a survival rate of 92% at 1year post-surgery and 88% at 3 years post-surgery. Vickers et al. attributed the relatively poor outcomes to the often-complicated health status of the patients’ middle ears. Recently, Knolke et al. (2025) reported data from all 79 patients operated in Hannover, where the FMT (the actuator of the VSB) was directly coupled to the round window of the cochlea. It was a retrospective study. A total of 16 revision surgeries were required of which 9 explantations/reimplantations (11%).

Skin problems around percutaneous implants may occur, these problems are classified according the Holgers scores: a Holgers score ≥ 2 usually leads to treatment. For a Holgers score of 2 (red and moist tissue), topical antibiotics. In addition, for a Holgers score of 3 and 4 was found (tissue granulations, less or more severe) systemic antibiotics. In the studies selected by Lagerkvist et al. one out of seven patients had once a Holgers score > 2, which required treatment. The average follow-up was 1.5 years. Teunissen, et al. (2024) also reported Holgers scores. The condition of the skin around the implant was systematically assessed at 1-3 months after surgery and annually thereafter, as well as during unplanned visits, due to some problem. A total of 5188 observations were available, i.e. a Holgers score of 2 was found in 6.1% of the observations and Holgers scores 3 and 4 together in 3.1% of the visits.

The implant survival rate for pBCD implants is favourable, not worse than that of other implantable devices (Bonebridge, VSB), but there may be problems with the skin around the percutaneous implant. Due to complicated and/or unstable medical conditions of the middle ear, the loss of VSB implants is relatively high; therefore, a BCD may be a safer choice (especially in chronic ear infections).

5.2 Recent developments in surgery

Apparently, the least complicated surgery is the placement of a percutaneous implant, while the most complex surgery is VSB implantation, requiring micro-surgery.

As for percutaneous implants, one of the latest, important surgical developments is the ‘minimal invasive Ponto surgery’ or MIPS, introduced in 2015 (Johansson and Holmberg, 2015). MIPS is based on punching the skin instead of making an incision, which was the standard surgical approach (the ‘linear incision procedure’).  In 2025, MIPS was updated; a new surgical set was introduced (Ganeyev et al., 2025), comprising a one-step drilling procedure (instead of three) with a novel drill design that provides more efficient bone removal and cooling. This simplification minimized discomfort for the patient. A major advantage of MIPS is the duration of the treatment, approx. 10 minutes, which is significantly shorter than that of the previous standard linear incision procedure (e.g. Calon et al., 2018; Kim et al., 2019) as well as that of surgical procedures to implant active tBCDs and the VSB. Due to the shorter drilling sequence, MIPS is particular suitable for local anaesthesia and implantation in the outpatient clinic. Nowadays (2025), MIPS is increasingly used outside the operation room (OR; see e.g., Sardiwalla et al., 2017; King et al., 2022; Cruz et al., 2023; Hyppolito et al., 2025). Sardiwalla et al. showed that the total cost of staff for MIPS in the outpatient clinic was 20 times lower than that for the linear incision procedure (standard performed in the OR), due to the reduced surgical time while avoiding the additional time to fulfil all the OR protocols. For the patient, MIPS in the outpatient clinic is experienced more as a treatment than as an operation and that is unique. MIPS is recommended for adults, not for children, amongst others, because of a thin skull. In terms of implant stability and skin problems, no significant differences were found between both surgical approaches (MIPS and linear incision procedure), for example by Calon et al. (2018) and Strijbos et al. (2021), both studied concerned multicentre randomized controlled trials.

5.3 MRI compatibility

An important point to consider when choosing an implantable device is MRI safety. MRI is a powerful and increasingly popular diagnostic tool. Statistics show that the incidence of MRI scans has been steadily increasing over the years. In 2020, the incidence was 66 MRI scans per 1000 habitants in the Netherlands.

Doshi et al. (2014) reviewed the literature and manufacturer information on MRI compatibility of BCDs and MEIs. Nowadays (2025) most implants are MRI safe up to at least 1.5T. It should be noted that with each upgrade, MRI safety is specified. Perhaps even more important is the degree of distortion of the MRI scan caused by the implanted device. The percutaneous implant affects the MRI image only locally (radius of 15 mm) while the active tBCDs, comprising a magnetic actuator, significantly distort the MRI scans, with a radius of 10 to 15 cm (Doshi et al., 2014). The best active tBCD with minimal distortion of the MRI image is the Osia device due to its piezoelectric actuator (Nassiri et al., 2024).

5.4 Cosmesis

About cosmesis, Krijnen et al. (2025) inquired individuals about the appearance of current sound processors of pBCDs and active tBCDs. They used several sets of photos of the latest generations of processors, worn by patients. About half of the 207 individuals were BCD users (mainly pBCD) and the other half were individuals with no experience with BCDs. The appearance of the pBCD sound processors was preferred by both groups. In addition, most individuals (73%) considered better hearing more important than having an attractive device. It was concluded that functionality seemed more important than the appearance of the device.

One reason affecting this outcome might be the size of the processor/device, which depends, amongst others, on the size of the battery. The battery consumption of the active tBCDs is relatively high, amongst others, due to the induction link. For example, battery life of the Osia device is only 20-30 hours (Cochlear 2023) despite the used powerful 675 battery. The other active tBCDs, also equipped with the 675 batteries, are less powerful (see Chapter 4), resulting in a longer battery life (up to 60-70 hours). Remarkably, according to the datasheets, the current consumption of the Ponto 5, a standard pBCD (Oticon 2021), is about four times lower than that of the Sentio device (Oticon 2024), while their capacities are similar (see Table 4.1). Because the current consumption is relatively low, most pBCDs use the (less powerful but significant smaller) 312 batteries. This, together with the relatively large coupling magnet and transmitting coil affects the size of the processor of the active tBCDs.

The appearance of the devices/processors is of secondary importance.

In summary, based on the subsections of this chapter 5, pBCDs are/remain an important alternative due to a relatively high implant stability, the least effect on MRI scans and the brief implant procedure (MIPS) that can be performed in outpatient clinics (only in adults).   

Chapter 1. General introduction; outdated fitting protocols?

This website deals with hearing solutions for patients with conductive or mixed hearing loss; if reconstructive surgery is not an option, there are several amplification options (hearing devices) available. In addition to the conventional behind-the-ear device (BTE), bone-conduction devices (BCDs) and middle-ear-implants (MEIs) can be used.

The website is designed in such a way that it can be used for educational purposes. Thereto, it starts with the basic audiological properties of the three types of amplification options, with the emphasis on their audiological capacities (output, gain) and speech recognition. It is noteworthy that BCDs and MEIs, unlike BTEs, have a limited audiological capacity. Therefore, it is challenging to find the best device for a particular patient, especially since the manufacturer’s brochures, are often quite optimistic. While audiological capacity is of high importance, other factor can also play a role in the choice of a specific device, such as safety and implant stability, aesthetics, medical issues, costs, and so on. These issues are discussed; however, surgical and economic issues will be just touched upon.

Chapters 2 to 4 cover the audiological capacities, resulting in a categorization of the available devices, and related fitting issues. Special attention is paid to declining effectiveness of these devices in the elderly due to an age-related progressive hearing loss. Chapters 5 addresses medical issues (post-surgical complications, implant loss). Chapter 6-8 cover bilateral applications, applications in children and the use of BCDs and MEIs in patients with sensorineural hearing loss if BTEs are contra-indicated (e.g., patients with chronic external otitis). Chapter 9 deals with borderline cases, including, amongst others, when you need to switch to a cochlear implant (CI).

The information provided may not be entirely accurate due to ongoing technological and surgical developments in this area, although, the website is updated regularly. Last update: chapters 1-4, June 2026 (see also Appendix 1: The history of the website).

1.1 Challenges and limitations of auditory implants for conductive and mixed hearing loss

A conductive hearing loss (component) is caused by some medical or anatomical middle ear problem or atresia of the ear canal. Surgical reconstruction is the preferred treatment. However, surgical treatment doesn’t always result in closing of the gap between the air- and bone-conduction thresholds (referred to as the air-bone gap or AB gap, e.g., Nadaraja et al., 2013, Shannon et al., 2023). In the search for what constitutes an ‘acceptable remaining hearing loss’, it has been argued that hearing thresholds of 30 dB HL or better are socially adequate (e.g. Browning, 1993; Persson et al., 2007). However, harmful effects of semi-permanent mild conductive hearing loss in children have been reported (Mondelli et al., 2010). Based on a review of the literature it has been stated that as little as 15 dB HL is a disabling hearing loss for children (see Northern & Downs, 2013, Chapter 1). This suggests that for children, the goal of intervention should be hearing thresholds around 15 dB HL (see also Chapter 7).

Today, there are several different types of hearing devices on the market for the rehabilitation of conductive and mixed hearing loss. Besides BTEs, these devices essentially directly stimulate the cochlea (BCDs and MEI with its actuator directly coupled to one of the cochlear windows), bypassing the impaired middle ear. Under certain circumstances such devices might be a better option than conventional BTEs, see Chapter 2.

Prescription of gain and output. What do we know about fitting (any type of) hearing devices in conductive and mixed hearing loss? What are desired values for gain and maximum output? For patients with pure sensorineural hearing loss, several well-validated fitting procedures are available, like the NAL-NL rule and the DSL i/o rule (Dillon, 2012). In such patients, hearing thresholds are elevated but not the loudness discomfort levels (or the highest sound pressure levels tolerated by the patient, referred to as LDLs), leading to a reduced dynamic range of hearing. Therefore, the output of BTEs has to be limited to prevent loud sounds from exceeding the patients’ LDLs.

In purely conductive hearing loss, not only the (air-conduction) hearing thresholds are elevated but also the LDLs, therefor the dynamic range of hearing is not affected. Theoretically, in pure conductive loss, the desired gain (amplification) should be equal to the width of the AB gap, resulting in normal input to the cochlea. In mixed hearing loss, additionally, about half of the sensorineural hearing loss component should be ‘compensated’, according to the NAL or DSL rule. However, the problem of relatively low capacity of BCDs and MEIs limits both the feasible gain and output, as discussed in the next chapters.

A univocal fitting procedure in which the limited capacity of these devices is considered, is missing. In order to discuss fitting issues (Chapter 4), we will start with discussing in detail the auditory capacities of the amplification options (Chapter 2) and how the limited capacities affect the gain setting of the devices, as chosen by the patients (Chapter 3).

References and Abbreviations

Abbreviations

‘AB-gap’: gap between the air- and bone-conduction thresholds

AI: Audibility Index

‘BCD’: Bone-conduction device (general term)

‘iBCD’: implantable ‘BCD’

‘pBCD’: percutaneous ‘BCD’

‘tBCD’: transcutaneous ‘BCD’

BTE: Behind-The-Ear device (or air-conduction device)

CHL: conductive hearing loss

CI: Cochlear Implant

CSOM: Chronic Supprative Otitis Media 

DR: dynamic range of hearing

DSL rule: hearing-aid fitting-rule developed by Scollie et al (2005)

EG: Effective Gain

FG: Functional Gain

FMT: ‘Floating-Mass’ Transducer

ILD: Interaural Loudness Difference

ITD: Interaural Time Difference

LDL: Loudness-Discomfort-Level

MCL: Most Comfortable Listening level

MEI: Middle Ear Implant

MIPS: Minimal Invasive Ponto Surgery

MPO: Maximum (Power) Output

MPS: phoneme scores using monosyllables

MRI: Magnetic Resonance Imaging

NAL rule: hearing-aid fitting-rule developed by Dillon and co-workers (Dillon 2012)

OME: Otitis Media with Effusion

OR: Operation Room

PTA: Pure Tone Average

PTA3: mean hearing lthresholds at 0.5, 1 and 2 kHz

PTA4: Average Threshold at 0.5, 1, 2, 4 kHz

PTAbc Pure Tone Average for bone-conduction

s.d.: standard deviation

SII: speech intelligibility index

SNHLc: Sensorineural Hearing Loss component

SRT: speech reception threshold

VSB: VibrantSoundbridge

WDRC: wide-band dynamic-range compression

WRS: word recognition score

References

Aarnisalo AA, Vasama JP, Hopsu E, Ramsay H. Long-term hearing results after stapes surgery: a 20-year follow-up. Otol Neurotol. 2003;24:567-71.

Agterberg MJ, Hol MK, Cremers CW, Mylanus EA, van Opstal J, Snik AF. Conductive hearing loss and bone conduction devices: restored binaural hearing? Adv Otorhinolaryngol. 2011;71:84-91

Agterberg MJ, Frenzel H, Wollenberg B, Somers T, Cremers CW, Snik AF. Amplification options in unilateral aural atresia: an active middle ear implant or a bone conduction device? Otol Neurotol. 2014;35:129-35.

Agterberg MJ, Snik AF, Hol MK, Van Wanrooij MM, Van Opstal AJ. Contribution of monaural and binaural cues to sound localization in listeners with acquired unilateral conductive hearing loss: improved directional hearing with a bone-conduction device. Hear Res. 2012;286:9-18.

Ahn J, Ryu G, Kang M, Cho YS.  Long-term Hearing Outcome of Canaloplasty With Partial Ossicular Replacement in Congenital Aural Atresia. Otol Neurotol. 2018 Jun;39(5):602-608.

Alomari HM. Binaural hearing with bone conduction stimulation. Eprints.soton.ac.uk/370832/1/Hala%20Alomari%20thesis.pdf. Thesis, 2014, University of Southampton

Amlani AM, Punch JL, Ching TY. Methods and applications of the audibility index in hearing aid selection and fitting. Trends Amplif. 2002;6(3):81-129.

Barbara M, Filippi C, Covelli E, Volpini L, Monini S. Ten years of active middle ear implantation for sensorineural hearing loss. Acta Otolaryngol. 2018a;138(9):807-814.

Barbara M, Volpini L, Filippi C, Atturo F, Monini S. A new semi-implantable middle ear implant for sensorineural hearing loss: three-years follow-up in a pilot patient’s group. Acta Otolaryngol. 2018b;138(1):31-35.

Baumgartner WD, Böheim K, Hagen R, Müller J, Lenarz T, Reiss S, Schlögel M, Mlynski R, Mojallal H, Colletti V, Opie J. The vibrant soundbridge for conductive and mixed hearing losses: European multicenter study results. Adv Otorhinolaryngol. 2010;69:38-50.

Baumgartner WD et al. A new transcutaneous bone conduction hearing implant: short term safety and efficacy in children. Otol Neurotol 2016;37:713-720.

Beck DL, Sockalingam R. Patient care and hearing instrument technology: tech topic. Hearing review 2010:17(4);44-47

Beleites T, Neudert M, Beutner D, Hüttenbrink KB, Zahnert T. Experience with vibroplasty couplers at the stapes head and footplate. Otol Neurotol. 2011;32:1468-72.

Beltrame AM, Martini A, Prosser S, Giarbini N, Streitberger C. Coupling the Vibrant Soundbridge to cochlea round window: auditory results in patients with mixed hearing loss. Otol Neurotol. 2009;30:194-201.

Bernardeschi D, Hoffman C, Benchaa T, Labassi S, Beliaeff M, Sterkers O, Grayeli AB.  Functional results of  Vibrant Soundbridge middle ear implants in conductive and mixed hearing losses. Audiol Neurootol. 2011;16:381-7.

Bezdjian A, Bruijnzeel H, Daniel SJ, Grolman W, Thomeer HGXM. Preliminary audiologic and peri-operative outcomes of the Sophono™ transcutaneous bone conduction device: A systematic review. Int J Pediatr Otorhinolaryngol. 2017;101:196-203.

Bosman AJ, Snik AFM, Van der Pouw CTM, Mylanus EAM, Cremers CWRJ: Audiometric evaluation of bilaterally fitted bone-anchored hearing aids. Audiol 2001;40:158-167.

Bosman AJ, Kruyt IJ, Mylanus EAM, Hol MKS, Snik AFM. Evaluation of an abutment-level superpower sound processor for bone-anchored hearing. Clin Otolaryngol. 2018a Feb 16.

Bosman AJ, Kruyt IJ, Mylanus EAM, Hol MKS, Snik AFM. On the evaluation of a superpower sound processor for bone-anchored hearing. Clin Otolaryngol. 2018b Apr;43(2):450-455.

Bosman AJ, Snik AFM, Mylanus EAM, Cremers CWRJ. Fitting range of the BAHA Cordelle. Int J Audiol 2006;45:429-37.

Bosman AJ, Snik AF, Hol MK, Mylanus EA. Evaluation of a new powerful bone-anchored hearing system: a comparison study. J Am Acad Audiol. 2013;24(6):505-13.

Bouhabel S, Arcand P, Saliba I. Congenital aural atresia: bone-anchored hearing aid vs. external auditory canal reconstruction. Int J Pediatr Otorhinolaryngol. 2012;76(2):272-7.

Boymans M, Goverts ST, Kramer SE, Festen JM, Dreschler WA. A prospective multi-centre study of the benefits of bilateral hearing aids. Ear Hear. 2008;29(6):930-41.

Bravo-Torres S, Der-Mussa C, Fuentes-López E. Active transcutaneous bone conduction implant: audiological results in paediatric patients with bilateral microtia associated with external auditory canal atresia. Int J Audiol. 2018 Jan;57(1):53-60.

Briggs R, Van Hasselt A, Luntz M, Goycoolea M, Wigren S, Weber P, Smeds H, Flynn M, Cowan R. Clinical Performance of a New Magnetic Bone Conduction Hearing  Implant System: Results from a Prospective, Multicenter, Clinical Investigation. Otol Neurotol. 2015 Jan 28

Brkic FF, Baumgartner WD, Riss D, Thurner T, Liu DT, Gstöttner W, Vyskocil E. progressive Sensorineural Hearing Loss in Vibrant Soundbridge Users Requiring Cochlear Implantation. J Pers Med. 2022 Jan 31;12(2):191.

Brkic FF, Riss D, Auinger A, Zoerner B, Arnoldner C, Baumgartner WD, Gstoettner W, Vyskocil E. Long-Term Outcome of Hearing Rehabilitation With An Active Middle Ear Implant. Laryngoscope. 2019;129:477-481.

Brkic FF, Riss D, Scheuba K, Arnoldner C, Gstöttner W, Baumgartner WD, Vyskocil E. Medical, Technical and Audiological Outcomes of Hearing Rehabilitation with the Bonebridge Transcutaneous Bone-Conduction Implant: A Single-Center Experience. J Clin Med. 2019 Oct 3;8(10):1614.

Browning GG. Reporting the benefits from middle ear surgery using the Glasgow Benefit Plot. Am J Otol. 1993;14(2):135-40.

Busch S, Lenarz T, Maier H. Comparison of Alternative Coupling Methods of the Vibrant Soundbridge Floating Mass Transducer. Audiol Neurootol. 2016;21(6):347-355.

Busch S, Kruck S, Spickers D, Leuwer R, Hoth S, Praetorius M, Plinkert PK, Mojallal H, Schwab B, Maier H, Lenarz T. First clinical experiences with a direct acoustic cochlear stimulator in comparison to preoperative fitted conventional hearing aids. Otol Neurotol. 2013;34(9):1711-8.

Busch S, Lenarz T, Maier H. Comparison of alternative coupling methods of the Vibrant Soundbridge floating mass transducer. Audiol Neurotol 2016; 21: 347-355

Burrell SP, Cooper HC, Proops DW. The bone anchored hearing aid–the third option for otosclerosis. J Laryngol Otol Suppl. 1996;21:31-7.

Calon TGA, Johansson ML, de Bruijn AJG, van den Berge H, Wagenaar M, Eichhorn E, Janssen MML, Hof JR, Brunings JW, Joore MA, Jonhede S, van Tongeren J, Holmberg M, Stokroos RJ. Minimally Invasive Ponto Surgery Versus the Linear Incision Technique with Soft Tissue Preservation for Bone Conduction Hearing Implants: A Multicenter Randomized Controlled Trial. Otol Neurotol. 2018 Aug;39(7):882-893.

Carlsson PU, Håkansson BE. The bone-anchored hearing aid: reference quantities and functional gain. Ear Hear 1997;18:34-41.

Christensen L, Smith-Olinde L, Kimberlain J, Richter GT, Dornhoffer JL. Comparison of traditional bone-conduction hearing AIDS with the Baha system. J Am Acad Audiol. 2010;21(4):267-73.

Claros P, Pujol Mdel C. Active middle ear implants: Vibroplasty? in children and adolescents with acquired or congenital middle ear disorders. Acta Otolaryngol. 2013;133(6):612-9.

Cochlear. Datasheet Osia, 2023. OSI144 Osia System ft OSI300 Datasheet ISS1.pdf

Colletti V, Soli SD, Carner M, Colletti L. Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window. Int J Audiol. 2006;45:600-8

Colletti L, Carner M, Mandalà M, Veronese S, Colletti V. The floating mass transducer for external auditory canal and middle ear malformations. Otol Neurotol. 2011;32(1):108-15.

Colletti V, Mandalà M, Colletti L. Electrocochleography in round window Vibrant Soundbridge implantation. Otolaryngol Head Neck Surg. 2012;146(4):633-40.

Colletti L, Mandala M, Colletti V. Long-term outcome of round window Vibrant Soundbridge implantation in extensive ossicular chain defects. Otolaryngol Head Neck Surg. 2015; 149:134-141

Cremers CW, Snik FM, Beynon AJ. Hearing with the bone-anchored hearing aid (BAHA, HC 200) compared to a conventional bone-conduction hearing aid. Clin Otolaryngol Allied Sci. 1992;17:275-9.

Cruz LDS, Danieli F, Håkansson MÅ, Johansson ML, Dos Santos FR, Mirândola Barbosa Reis AC, Hyppolito MA. Minimally invasive surgery as a new clinical standard for bone anchored hearing implants-real-world data from 10 years of follow-up and 228 surgeries. Front Surg. 2023 Jul 3;10:1209927.

Danhauer JL, Johnson CE, Mixon M. Does the evidence support use of the Baha implant system (Baha) in patients with congenital unilateral aural atresia? J Am Acad Audiol. 2010;21:274-86.

Daniels N, Sabin J. The ethics of accountability in managed care reform. Health Aff (Millwood). 1998 Sep-Oct;17(5):50-64.

Debeaupte M, Decullier E, Tringali S, Devèze A, Mom T, Darrouzet V, Truy E. Evolution of the reliability of the fully implantable middle ear transducer over successive generations. Otol Neurotol. 2015;36:625-30.

Declau F, Cremers C, Van de Heyning. Diagnosis and management strategies in congenital atresia of the external auditory canal. Br J Audiol 1999; 33:313-327

Desmet JB, Bosman AJ, Snik AF, Lambrechts P, Hol MK, Mylanus EA, De Bodt M, Van de Heyning P. Comparison of sound processing strategies for osseointegrated bone conduction implants in mixed hearing loss: multiple-channel nonlinear versus single-channel linear processing. Otol Neurotol. 2013;34:598-603.

den Besten CA,  Harterink E,  McDermott AL, Hol MK. Clinical results of Cochlear™ BIA300 in children: Experience in two tertiary referral centers.  “International journal of pediatric otorhinolaryngology.” Int J Pediatr Otorhinolaryngol. 2015;79:2050-2055

den Besten CA, Monksfield P, Bosman A, Skarzynski PH, Green K, Runge C, Wigren S, Blechert JI, Flynn MC, Mylanus EAM, Hol MKS. Audiological and clinical outcomes of a transcutaneous bone conduction hearing implant: Six-month results from a multicentre study. Clin Otolaryngol. 2019;44(2):144-157.

den Besten CA, Vogt K, Bosman AJ, Snik AFM, Hol MKS, Agterberg MJH. The Merits of Bilateral Application of Bone-Conduction Devices in Children With Bilateral Conductive Hearing Loss. Ear Hear. 2020;41(5):1327-1332.

Denoyelle F, Coudert C, Thierry B, Parodi M, Mazzaschi O, Vicaut E, Tessier N, Loundon N, Garabedian EN. Hearing rehabilitation with the closed skin bone-anchored implant Sophono Alpha1: results of a prospective study in 15 children with ear atresia. Int J Pediatr Otorhinolaryngol. 2015;79(3):382-7.

de Wolf MJ, Hendrix S, Cremers CW, Snik AF. Better performance with bone-anchored hearing aid than acoustic devices in patients with severe air-bone gap. Laryngoscope 2011;121:613-6.

de Wolf MJ, Hol MK, Huygen PL, Mylanus EA, Cremers CW. Nijmegen results with application of a bone-anchored hearing aid in children: simplified surgical technique. Ann Otol Rhinol Laryngol. 2008;117(11):805-14.

Dillon H, Storey L. The National Acoustic Laboratories’ procedure for selecting the saturation sound pressure level of hearing aids: theoretical derivation. Ear Hear 1998;19:255-66.

Dillon H. 2012. Prescribing hearing aid performance. Chapter 10 in ‘Hearing aids’, Dillon, 2012. Thieme Verlag, Stuttgart, Germany

Dimitriadis PA, Farr MR, Allam A, Ray J. Three year experience with the cochlear BAHA attract implant: a systematic review of the literature. BMC Ear Nose Throat Disord. 2016 Oct 1;16:12.

Dimitriadis PA, Hind D, Wright K, Proctor V, Greenwood L, Carrick S, Ray J. Single-center Experience of Over a Hundred Implantations of a Transcutaneous Bone Conduction Device. Otol Neurotol. 2017;38:1301-1307.

Doshi J, Schneiders S, Foster K, Reid A, McDermott AL. Magnetic resonance imaging and bone anchored hearing implants: pediatric considerations. Int J Pediatr Otorhinolaryngol. 2014;78:277-9.

Dumon T, Gratacap B, Firmin F, Vincent R, Pialoux R, Casse B, Firmin B. Vibrant Soundbridge middle ear implant in mixed hearing loss. Indications, techniques, results. Rev Laryngol Otol Rhinol (Bord). 2009;130:75-81.

Dun CA, Agterberg MJ, Cremers CW, Hol MK, Snik AF. Bilateral bone conduction devices: improved hearing ability in children with bilateral conductive hearing loss. Ear Hear. 2013;34:806-8.

Dun CA, Faber HT, de Wolf MJ, Mylanus EA, Cremers CW, Hol MK. Assessment of more than 1,000 implanted percutaneous bone conduction devices: skin reactions and implant survival. Otol Neurotol 2012;33:192-8.

Edfelt L, Stromback K, Grendin J et al., Evaluation of the cost-utility in middle ear implantation in the ‘Nordic’ school; a multi-center study in Sweden and Norway. Acta Otolaryngol. 2014;134: 19-25.

Ellsperman SE, Nairn EM, Stucken EZ. Review of Bone Conduction Hearing Devices. Audiol Res. 2021;11(2):207-219.

Ernst A, Todt I, Wagner J. Safety and effectiveness of the Vibrant Soundbridge in treating conductive and mixed hearing loss: A systematic review. Laryngoscope. 2016;126:1451-7.

Evans AK, Kazahaya K. Canal atresia: “surgery or implantable hearing devices? The expert’s question is revisited”. Int J Pediatr Otorhinolaryngol. 2007;71:367-74

Fan X, Yang T, Niu X, Wang Y, Fan Y, Chen X. Long-term Outcomes of Bone Conduction Hearing Implants in Patients with Bilateral Microtia-atresia. Otol Neurotol. 2019 Sep;40(8):998-1005.

Fan X, Yang T, Niu X, Wang Y, Fan Y, Chen X. Long-term Outcomes of Bone Conduction Hearing Implants in Patients With Bilateral Microtia-atresia. Otol Neurotol. 2019 Sep;40(8):998-1005.

Farnoosh S, Mitsinikos FT, Maceri D, Don DM. Bone-Anchored Hearing Aid vs. Reconstruction of the External Auditory Canal in Children and Adolescents with Congenital Aural Atresia: A Comparison Study of Outcomes.Front Pediatr. 2014,22;2:5.

Favoreel A, Heuninck E, Mansbach AL. Audiological benefit and subjective satisfaction of children with the ADHEAR audio processor and adhesive adapter. Int J Pediatr Otorhinolaryngol. 2020;129:109729.

Flynn MC, Hedin A, Halvarsson G, Good T, Sadeghi A. Hearing performance benefits of a programmable power baha® sound processor with a directional microphone for patients with a mixed hearing loss. Clin Exp Otorhinolaryngol. 2012;5 Suppl 1:S76-81.

Flynn MC, Sadeghi A, Halvarsson G. Baha solutions for patients with severe mixed hearing loss. Cochlear Implants Int. 2009;10 Suppl 1:43-7.

Frenzel H, Hanke F, Beltrame M, Wollenberg B. Application of the Vibrant Soundbridge in bilateral congenital atresia in toddlers. Acta Otolaryngol. 2010 ;130(8):966-70.

Frenzel H, Sprinzl G, Streitberger C, Stark T, Wollenberg B, Wolf-Magele A, Giarbini N, Strenger T, Muller J, Hempel JM. The Vibrant Soundbridge in Children and Adolescents: Preliminary European Multicenter Results.  Otol Neurotol. 2015;36(7):1216-1222.

Fuchsmann C, Tringali S, Disant F, Buiret G, Dubreuil C, Froehlich P, Truy E. Hearing rehabilitation in congenital aural atresia using the bone-anchored hearing aid: audiological and satisfaction results. Acta Otolaryngol. 2010;130(12):1343-51.

Ganeyev M, Shah FA, Thomsen P, Palmquist A, Johansson ML. Mechanical and thermal efficiency of a single drill system for bone-anchored hearing implants. PLoS One. 2025 May 30;20(5):e0311026.

Gantner S, Epp A, Pollotzek M, Hempel JM. Long-term results and quality of life after vibrant soundbridge implantation (VSBs) in children and adults with aural atresia. Eur Arch Otorhinolaryngol. 2024;281(1):129-139.

Garin P, Schmerber S, Magnan J, Truy E, Uziel A, Triglia JM, Bebear JP, Labassi S, Lavieille JP. Bilateral vibrant soundbridge implantation: audiologic and subjective benefits in quiet and noisy environments. Acta Otolaryngol. 2010;130(12):1370-8.

Gascon A, Bagatto M, Scollie S, Cowan C, Hodgetts W. Clinical evaluation of DSL-BCD fittings: assessing output-to-targets for percutaneous bone conduction devices in practice. Front. Audiol. Otol. 3:1714575.

Gawliczek T, Munzinger F, Anschuetz L, Caversaccio M, Kompis M, Wimmer W. Unilateral and Bilateral Audiological Benefit With an Adhesively Attached, Noninvasive Bone Conduction Hearing System. Otol Neurotol. 2018;39(8):1025-1030.

Gerdes T, Salcher RB, Schwab B, Lenarz T, Maier H. Comparison of Audiological Results Between a Transcutaneous and a Percutaneous Bone Conduction Instrument in Conductive Hearing Loss. Otol Neurotol. 2016;37:685-91.

Ghoncheh M, Busch S, Lenarz T, Maier H. A Novel Method to Determine the Maximum Output of Individual Patients for an Active Transcutaneous Bone Conduction Implant Using Clinical Routine Data. Ear Hear. 2024;45(1):219-226.

Goode RI. Current status and future of implantable electromagnetic hearing aids. Otolaryngol Clin N Am. 1995;28:141-146

Gordon K, Henkin Y, Kral A. Asymmetric Hearing During Development: The Aural Preference Syndrome and Treatment Options. Pediatrics. 2015;136(1):141-53.

Groth J, Sondergaard MB. Disturbance caused by varying propagation delays in non-occluding hearing aid fittings. Int J Audiol. 2004;43:594-599

Gunduz B, Atas A, Bayazıt YA, Goksu N, Gokdogan C, Tutar H. Functional outcomes of Vibrant Soundbridge applied on the middle ear windows in comparison with conventional hearing aids. Acta Otolaryngol. 2012;132:1306-10.

Håkansson B, Tjellstrom A, Rosenhall U. Hearing thresholds with direct bone conduction versus conventional bone conduction. Scand Audiol 1984;13:3-13.

Hol MKS, Cremers CWRJ, Coppens-Schellekens W, Snik AFM. The BAHA Softband. A new treatment for young children with bilateral congenital aural atresia. Int J Ped Otorhinolaryngol 2005; 69:973-980

Hol MK, Spath MA, Krabbe PF, van der Pouw CT, Snik AF, Cremers CW, Mylanus EA.  The bone-anchored hearing aid: quality-of-life assessment. Arch Otolaryngol Head Neck Surg. 2004;130:394-9.

Hol MKS, Nelissen R, Agterberg M, et al. Comparison between a new implantable transcutaneous bone conductor and percutaneous bone-conduction hearing implant. Otol Neurotol 2013;34:1071-5.

Hol MKS, Aukema TW, Ubbink SWJ, Lindholm D, Holmberg M, Arndt S, Wolf S, Monksfield P, Marwa M, Lenarz T, Busch S, Tysome J, Asher A, Kok H, Mylanus EM. A Prospective, Multicenter Clinical Investigation on the Safety and Performance of a New Active transcutaneous Bone-Anchored Implant System. Otol Neurotol. 2026;47(2):287-296.

Holland Brown T, Fitzgerald O’Connor I, Bewick J, Morley C. Bone conduction hearing kit for children with glue ear. BMJ Innovations. 2021:7:600-3. 

Holland Brown T, Salorio-Corbetto M, Gray R, James Best A, Marriage JE. Using a Bone-Conduction Headset to Improve Speech Discrimination in Children With Otitis Media With Effusion. Trends Hear. 2019 Jan-Dec;23:2331216519858303.

Huber AM, Sim JH, Xie YZ, Chatzimichalis M, Ullrich O, Röösli C. The Bonebridge: preclinical evaluation of a new transcutaneously-activated bone anchored hearing device. Hear Res 2013;301:93-9.

Huinck WJ, Mylanus EAM, Snik AFM. Expanding unilateral cochlear implantation criteria for adults with bilateral acquired severe sensorineural hearing loss. Eur Arch Otorhinolaryngol. 2019 May;276(5):1313-1320.

Hultcrantz M, Lanis A. A five-year follow-up on the osseointegration of bone-anchored hearing device implantation without tissue reduction. Otol Neurotol. 2014;35:1480-5.

Humes  LE. Dimensions of hearing aid outcomes. J Am Acad Audiol. 1999; 10:26-39

Hüttenbrink KB, Zahnert T, Bornitz M, Beutner D. TORP-vibroplasty: a new alternative for the chronically disabled middle ear. Otol Neurotol. 2008;29(7):965-71.

Hyppolito MA, Reis ACMB, de Oliveira LC, da Silva KE, Garcia DM, Danieli-Hyppolito F. Outcomes of a New Minimally Invasive Single-Drill Procedure for Installing Percutaneous Bone-Anchored Hearing Implants in Adult Patients. Otol Neurotol. 2025 Jun 1;46(5):e190-e197.

Ihler F, Blum J, Berger MU, Weiss BG, Welz C, Canis M. The Prediction of Speech Recognition in Noise With a Semi-Implantable Bone Conduction Hearing System by External Bone Conduction Stimulation With Headband: A Prospective Study. Trends Hear. 2016; 3;20. pii: 2331216516669330.

Iliadou V, Van Den Bogaert K, Eleftheriades N, Aperis G, Vanderstraeten K, Fransen E, Thys M, Grigoriadou M, Pampanos A, Economides J, Iliades T, Van Camp G, Petersen MB. Monogenic nonsyndromic otosclerosis: audiological and linkage analysis in a large Greek pedigree. Int J Pediatr Otorhinolaryngol. 2006;70(4):631-7.

Işeri M, Orhan KS, Kara A, Durgut M, Oztürk M, Topdağ M, Calışkan S. A new transcutaneous bone anchored hearing device – the Baha® Attract System: the first experience in Turkey. Kulak Burun Bogaz Ihtis Derg. 2014;24:59-64.

Iseri M, Orhan KS, Tuncer U, Kara A, Durgut M, Guldiken Y, Surmelioglu O. Transcutaneous Bone-Anchored Hearing Aids Versus Percutaneous Ones: Multicenter Comparative Clinical Study. Otol Neurotol. 2015 Feb 27. [Epub ahead of print]

ISO 7029, 2017. Acoustics. Statistical distribution of hearing thresholds related to age and genders. Third edition, 2017-01, International Organization for Standardization, Geneva, Switzerland

Janssen RM, Hong P, Chadha NK. Bilateral bone-anchored hearing aids for bilateral permanent conductive hearing loss: a systematic review. Otolaryngol Head Neck Surg. 2012;147(3):412-22.

Johansson M, Holmberg M. Design and clinical evaluation of MIPS—A new perspective on tissue preservation. White Pap Oticon Medical, Askim, Sweden 2015. October, Rep No M524252.

Kahue CN, Carlson ML, Daugherty JA, Haynes DS, Glasscock ME 3rd. Middle ear implants for rehabilitation of sensorineural hearing loss: a systematic review of FDA approved devices. Otol Neurotol. 2014;35(7):1228-37.

Killion MC, Christensen. The case of the missing dots: AI and SNR loss. Hear J 1998;41(5):1-6.

Killion MC, Mueller HG. Twenty years later, a new count-the-dots method. Hearing Journal 2010;63(1):11-17.

Kim HHS, Kari E, Copeland BJ, Moore GF, Shaia WT, Cosetti MK, Wazen JJ, Wanna GB, Chang CJ. Standardization of the Punch Technique for the Implantation of Bone Anchored Auditory  devices: Evaluation of the MIPS Surgical Set. Otol Neurotol. 2019 Jul;40(6):e631-e635.

King J, Leon I, Squires L. In-office bone-anchored hearing implants via minimally invasive punch technique in a veteran population. Otolaryngol Head Neck Surg 2022;167:959–63.

Knolke N, Busch S, Lenarz T, Naier H. Vibrant Soundbridge Round Window Vibroplasty: Safety, Coupling Efficiency, and Speech Outcome of the Most Common Coupling Modalities. Audiol Neurootol. 2025; 30:392-403

Koci V, Seebacher J, Weichbold V, Zorowka P, Wolf-Magele A, Sprinzl G, Stephan K. Improvement of sound source localization abilities in patients with bilaterally supplied active middle ear implants. Acta Otolaryngol 2016;136:692-698

Kontorinis G, Lenarz T, Mojallal H, Hinze AL, Schwab B. Power stapes: an alternative method for treating hearing loss in osteogenesis imperfecta? Otol Neurotol. 2011;32(4):589-95.

Kompis M, Kurz A, Pfiffner F, Senn P, Arnold A, Caversaccio M. Is complex signal processing for bone conduction hearing aids useful? Cochlear Implants Int. 2014;15 Suppl 1:S47-50.

Kompis M, Krebs M, Häusler R. Speech understanding in quiet and in noise with the bone-anchored hearing aids Baha Compact and Baha Divino. Acta Otolaryngol. 2007;127(8):829-35.

Krijnen HK, Aukema TW, Hol MKS. Participant valued appearance of bone conduction devices: a comparison between percutaneous and transcutaneous systems. Eur Arch Otorhinolaryngol. 2025 Sep;282(9):4467-4475.

Kruyt IJ, Monksfield P, Skarzynski PH, Green K, Runge C, Bosman A, Blechert JI, Wigren S, Mylanus EAM, Hol MKS. Results of a 2-Year Prospective Multicenter Study Evaluating Long-term Audiological and Clinical Outcomes of a Transcutaneous Implant for Bone Conduction Hearing. Otol Neurotol. 2020;41(7):901-911.

Kruyt IJ, Bakkum KHE, Caspers CJI, Hol MKS. The efficacy of bone-anchored hearing implant surgery in children: A systematic review. Int J Pediatr Otorhinolaryngol. 2020;132:109906.

Kulasegarah J, Burgess H, Neeff M, Brown CRS. Comparing audiological outcomes between the Bonebridge and bone conduction hearing aid on a hard test band: Our experience in children with atresia and microtia. Int J Pediatr Otorhinolaryngol. 2018 Apr;107:176-182.

Kunst SJ, Leijendeckers JM, Mylanus EA, Hol MK, Snik AF, Cremers CW. Bone-anchored hearing aid system application for unilateral congenital conductive hearing impairment: audiometric results. Otol Neurotol. 2008;29(1):2-7.

Kuppler K, Lewis M, Evans AK. A review of unilateral hearing loss and academic performance: is it time to reassess traditional dogmata? Int J Pediatr Otorhinolaryngol. 2013;77(5):617-22.

Kürsten R, Schneider B, Zrunek M. Long-term results after stapedectomy versus stapedotomy. Am J Otol. 1994;15:804-6.

Kurz A, Caversaccio M, Kompis M. Hearing performance with 2 different high-power sound processors for osseointegrated auditory implants. Otol Neurotol. 2013;34(4):604-10.

Kurz A, Flynn M, Caversaccio M, Kompis M. Speech understanding with a new implant technology: a comparative study with a new non-skin penetrating Baha system. Biomed Res Int. 2014;2014:416205.

Lagerkvist H, Carvalho K, Holmberg M, Petersson U, Cremers C, Hultcrantz M. Ten years of experience with the Ponto bone-anchored hearing system-A systematic literature review. Clin Otolaryngol. 2020;45:667-680.

Leinung M, Zaretsky E, Lange BP, Hoffmann V, Stöver T, Hey C. Vibrant Soundbridge® in preschool children with unilateral aural atresia: acceptance and benefit. Eur Arch Otorhinolaryngol. 2017;274:159-165.

Lenarz T, Verhaert N, Desloovere C, Desmet J, D’hondt C, González JC, Kludt E, Macías AR, Skarżyński H, Van de Heyning P, Vyncke C, Wasowski A. A comparative study on speech in noise understanding with a direct acoustic cochlear implant in subjects with severe to profound mixed hearing loss.Audiol Neurootol. 2014;19(3):164-74

Lenarz T, Zwartenkot JW, Stieger C, Schwab B, Mylanus EA, Caversaccio M, Kompis M, Snik AF, D’hondt C, Mojallal H. Multicenter study with a direct acoustic cochlear implant. Otol Neurotol. 2013 34:1215-25

Linder T, Schlegel C, DeMin N, van der Westhuizen S. Active middle ear implants in patients undergoing subtotal petrosectomy: new application for the Vibrant Soundbridge device and its implication for lateral cranium base surgery. Otol Neurotol. 2009;30(1):41-7.

Liu Y, Chen P, Yang L, Zhu J, Yang J, Wang D, Ren R, Li Y, Zhao C, Zhao S. Optimal Choice for Improving the Hearing in Children with Unilateral Microtia and Atresia: Softband or Adhesive Adapter? Audiol Neurootol. 2023;28(2):128-137.

Macnamara M, Phillips D, Proops DW. The bone anchored hearing aid (BAHA) in chronic suppurative otitis media (CSOM). J Laryngol Otol Suppl. 1996;21:38-40.

Magliulo G, Turchetta R, Iannella G, Valpega di Massino R, de Vincentiis M. Sophono Alpha system and subtotal petrosectomy with external auditory canal blind sac closure. Eur Arch Otorhinolaryngol 2015;272:2183-2190

Magnan J, Manrique M, Dillier N, Snik A, Hausler R. International consensus on middle ear implants. Acta Otolaryngol. 2005;125:920-921

Maier H. Improving the evidence base of audiological indication ranges from pre-clinical and clinical measurements. Keynote, Osseo 2023, Denver, US, 2023.

Maier H, Hinze AL, Gerdes T, Busch S, Salcher R, Schwab B, Lenarz T. Long-Term Results of Incus Vibroplasty in Patients with Moderate-to-Severe Sensorineural Hearing Loss. Audiol Neurootol. 2015;18;20:136-46.

Maier H. et al. Consensus Statement on Bone Conduction Devices and Active Middle Ear Implants in Conductive and Mixed Hearing Loss. Otol Neurotol. 2022;43(5):513-529

Mandala M, Colletti L, Colletti V. Treatment of the atretic ear with round window vibrant soundbridge implantation in infants and children: electrocochleography and audiologic outcomes. Otol Neurotol. 2011;32(8):1250-5.

Mancheño M, Aristegui M, Sañudo JR. Round and Oval Window Anatomic Variability: Its Implication for the Vibroplasty Technique. Otol Neurotol. 2017;38:e50-e57.

Marsella P, Scorpecci A, Pacifico C, Tieri L. Bone-anchored hearing aid (Baha) in patients with Treacher Collins syndrome: tips and pitfalls. Int J Pediatr Otorhinolaryngol. 2011;75(10):1308-12.

Marino R, Linton N, Eikelboom RH, Statham E, Rajan GP. A comparative study of hearing aids and round window application of the vibrant sound bridge (VSB) for patients with mixed or conductive hearing loss. Int J Audiol. 2013;52:209-18.

Marzo SJ, Sappington JM, Shohet JA. The Envoy Esteem implantable hearing system. Otolaryngol Clin North Am. 2014;47(6):941-52.

McDermott AL, Williams J, Kuo M, Reid A, Proops D. The Birmingham pediatric bone-anchored hearing aid program: a 15-year experience. Otol Neurotol. 2009;30(2):178-83.

Merkus P, van Loon MC, Smit CF, Smits C, de Cock AF, Hensen EF. Decision making in advanced otosclerosis: an evidence-based strategy. Laryngoscope. 2011;121:1935-41.

Mertens G, Desmet J, Snik, AFM, Van den Heyning P. An objective method to determine maximum output and dynamic range of an active bone conduction implant: The BonebridgeTM. Otol Neurotol 2014;35:1126-30

Mlynski R. The Vibrant Soundbridge, past, present and the future. Presentation at Otology Jubilee, May 9th, 2014, Halle (Saale) Germany

Mondelli MF, Carvalho FR, Feniman MR, Lauris JR. Mild hearing loss: performance in the Sustained Auditory Attention Ability Test. Pro Fono. 2010;22(3):245-50.

Mojallal H, Schwab B, Maier H, Lenarz T. First clinical experiences with a direct acoustic cochlear stimulator in comparison to preoperative fitted conventional hearing aids. Otol Neurotol. 2013;34:1711-8.

Mueller, G. & Killion, M. (1990). An Easy Method for Calculating the Articulation Index. The Hearing Journal 1990;43(9):14-7.

Mylanus EA, van der Pouw KC, Snik AF, Cremers CW. Intraindividual comparison of the bone-anchored hearing aid and air-conduction hearing aids. Arch Otolaryngol Head Neck Surg. 1998 Mar;124(3):271-6.

Nadaraja GS, Gurgel RK, Kim J, Chang KW. Hearing outcomes of atresia surgery versus osseointegrated bone conduction device in patients with congenital aural atresia: a systematic review. Otol Neurotol. 2013;34:1394-9

Nassiri AM, Messina SA, Benson JC, Lane JI, McGee KP, Trzasko JD, Carlson ML. Magnetic Resonance Imaging Artifact Associated With Transcutaneous Bone Conduction Implants: Cholesteatoma and Vestibular Schwannoma Surveillance. Otolaryngol Head Neck Surg. 2024;170:187-194.

Nelissen RC, Mylanus EA, Cremers CW, Hol MK, Snik AF. Long-term Compliance and Satisfaction With Percutaneous Bone Conduction Devices in Patients With Congenital Unilateral Conductive Hearing Loss. Otol Neurotol. 2015;36:826-33.

Nelissen RC, Stalfors J, de Wolf MJ, Flynn MC, Wigren S, Eeg-Olofsson M, Green K, Rothera MP, Mylanus EA, Hol MK. Long-term stability, survival, and tolerability of a novel osseointegrated implant for bone conduction hearing: 3-year data from a multicenter, randomized, controlled, clinical investigation. Otol Neurotol. 2014 Sep;35(8):1486-91.

Neumann K, Thomas JP, Voelter C, Dazert S. A new adhesive bone conduction hearing system effectively treats conductive hearing loss in children. Int JPediatr Otorhinolaryngol. 2019;122:117-125.

Northern JL & Downs MP. Hearing in children. 1991; Baltimore: William and Wilkins

Orús Dotú C, Santa Cruz Ruíz S, De Juan Beltrán J, Batuecas Caletrio A, Venegas   Pizarro Mdel P, Muñoz Herrera A. Treatment of severe to profound mixed hearing loss with the BAHA Cordelle II. Acta Otorrinolaringol Esp. 2011;62:205-12.

Osborne MS, Child-Hymas A, McDermott AL. Longitudinal study of use of the pressure free, adhesive bone conducting hearing system in children at a tertiary centre. Int J Pediatr Otorhinolaryngol. 2020;138:110307.

Oticon. Datasheet Sentio, 2024. 275144en_PI_Sentio 1 Mini_Version A_2024.04_Original file.pdf

Oticon. Datasheet Ponto 5 mini, 2021. 232708us_pi_ponto-5-mini_2021-05_low.pdt

Pelosi S, Carlson ML, Glasscock ME 3rd. Implantable hearing devices: the Ototronix MAXUM system. Otolaryngol Clin North Am. 2014;47(6):953-65.

Pfiffner F, Caversaccio MD, Kompis M. Comparisons of sound processors based on osseointegrated implants in patients with conductive or mixed hearing loss. Otol Neurotol. 2011;32(5):728-35.

Pfiffner F, Caversaccio MD, Kompis M. Audiological results with Baha in conductive and mixed hearing loss. Adv Otorhinolaryngol. 2011;71:73-83.

Plomp R. Auditory handicap of hearing impairment and the limited benefit of hearing aids. J Acoust Soc Am. 1978;63:533-549

Powell HRF, Rolfe AM, Birman CS. A comparative study of audiologic outcomes for two transcutaneous bone-anchored hearing devices. Otol Neurotol 2015;36:1525-1531.

Priwin C, Jönsson R, Hultcrantz M, Granström G. BAHA in children and adolescents with unilateral or bilateral conductive hearing loss: a study of outcome. Int J Pediatr Otorhinolaryngol. 2007;71(1):135-45.

Pulcherio JO, Bittencourt AG, Burke PR, Monsanto R, de Brito R, Tsuji RK, Bento RF. Carina® and Esteem®: a systematic review of fully implantable hearing devices. PLoS One. 2014;17;9(10)

Ramakrishnan Y, Davison T, Johnson IJM. How we do it: Softband management of glue ears. Clin Otolaryngol 2006;31:224-227.

Rameh C, Meller R, Lavieille JP, Deveze A, Magnan J. Long-term patient satisfaction with different middle ear hearing implants in sensorineural hearing loss. Otol Neurotol 2010;31:883-892

Ratuszniak A, Skarzynski PH, Gos E, Skarzynski H. The Bonebridge implant in older children and adolescents. Int J Pediatr Otorhinolaryngol. 2019 Mar;118:97-102.

Reinfeldt S, Hakansson B, Taghavi H, Eeg-Olofsson. New developments in bone-conduction hearing implants. Medical Devices: Evidence and Research 2015:8:79-93

Ren R, Zhao S, Wang D, Li Y, Ma X, Li Y, Fu X, Chen P, Dou J. Audiological effectiveness of Bonebridge implantation for bilateral congenital malformation of the external and middle ear. Eur Arch Otorhinolaryngol. 2019 Oct;276(10):2755-2762

Ricci G, Volpe AD, Faralli M, Longari F, Lancione C, Varricchio AM, Frenguelli A. Bone-anchored hearing aids (Baha) in congenital aural atresia: personal experience. Int J Otorhinolaryng Pediatr. 2011;75(3):342-6.

Roman S, Denoyelle F, Farinetti A, Garabedian EN, Triglia JM. Middle ear implant in conductive and mixed congenital hearing loss in children. Int J Pediatric Otorhinolaryngol 2012;76:1775-1778.

Roman S, Nicollas R, Triglia JM. Practice guidelines for bone-anchored hearing aids in children. Eur Ann Otorhinolaryngol Head Neck Dis. 2011;128(5):253-8.

Rotteveel LJ, Proops DW, Ramsden RT, Saeed SR, van Olphen AF, Mylanus EA. Cochlear implantation in 53 patients with otosclerosis: demographics, computed tomographic scanning, surgery, and complications. Otol Neurotol. 2004;25:943-52.

Rotteveel LJ, Snik AF, Cooper H, Mawman DJ, van Olphen AF, Mylanus EA. Speech perception after cochlear implantation in 53 patients with otosclerosis: multicentre results. Audiol Neurootol. 2010;15(2):128-36.

Sardiwalla Y, Jufas N, Morris DP. Direct cost comparison of minimally invasive punch technique versus traditional approaches for percutaneous bone anchored hearing devices. J Otolaryngol Head Neck Surg. 2017 Jun 12;46(1):46.

Schmerber S, Deguine O, Marx M, Van de Heyning P, Sterkers O, Mosnier I, Garin P, Godey B, Vincent C, Venail F, Mondain M, Deveze A, Lavieille JP, Karkas A. Safety and effectiveness of the Bonebridge transcutaneous active direct-drive bone-conduction hearing implant at 1-year device use. Eur Arch Otorhinolaryngol. 2016;30. [Epub ahead of print]

Schwab B, Kludt E, Maier H, Lenarz T. Subtotal petrpsectomy and Codacs: new possibilities in ears with chronic infection. Eur Arch Otorhinolaryngol 2016;273:1387-1391

Scollie S, Seewald R, Cornelisse L, Moodie S, Bagatto M, Laurnagaray D, Beaulac S, Pumford J. The Desired Sensation Level multistage input/output algorithm. Trends Amplif. 2005;9(4):159-97. 

Shannon CM, Gutierrez JA 3rd, Nguyen SA, Meyer TA, Lambert PR. Comparison of Outcomes of Surgery Versus Implantable Device for the Treatment of Hearing Loss Associated With Congenital Aural Atresia: A Systematic Review and Meta-Analysis. Otol Neurotol. 2023 Sep 1;44(8):758-766. 

Shimizu Y, Puria S, Goode RL. The floating mass transducer on the round window versus attachment to an ossicular replacement prosthesis. Otol Neurotol. 2011;32(1):98-103.

Shin JW, Kim SH, Choi JY, Park HJ, Lee SC, Choi JS, Park HQ, Lee HK. Surgical and Audiologic Comparison Between Sophono and Bone-Anchored Hearing Aids Implantation. Clin Exp Otorhinolaryngol. 2016;9:21-6.

Siegert R, Kanderske J. A new semi-implantable transcutaneous bone conduction device: clinical, surgical, and audiologic outcomes in patients with congenital ear canal atresia. Otol Neurotol 2013;34:927-34.

Skarzynski PH, Ratuszniak A, Osinska K, Koziel M, Krol B, Cywka KB, Skarzynski H. A Comparative Study of a Novel Adhesive Bone Conduction Device and Conventional Treatment Options for Conductive Hearing Loss. Otol Neurotol. 2019;40:858-864.

Skoda-Türk R, Welleschik B. Air conduction or bone conduction hearing aids? Comments on hearing aids for sound conduction deafness. Laryngol Rhinol Otol (Stuttg) 1981;60:478-83.

Souza MA, Vallejos Riart SL, Souza SR, de Brito R, Bento RF. Complications of Transcutaneous Protheses – A Systematic Review of Publications Over the Past 10 years. Int Arch Otorhinolaryngol. 2022 Feb 4;26(3):e505-e512.

Snik AFM, Cremers CWRJ. Vibrant semi implantable hearing device with digital sound processing. Effective gain and speech perception. Arch Otolaryngol Head Neck Surg 2001;127:1433-1437

Snik AF, Bosman AJ, Mylanus EA, Cremers CW. Candidacy for the bone-anchored hearing aid. Audiol Neurootol. 2004;9(4):190-6.

Snik A, Noten J, Cremers C. Gain and maximum output of two electromagnetic  middle ear implants: are real ear measurements helpful? J Am Acad Audiol. 2004;15(3):249-57.

Snik A, Hodgetts B, Mylanus E, Maier H, Snapp H, Kompis M, Agterberg M.  Percutaneous versus transcutaneous bone-conduction devices: part 1, audiological capacities. 2026 submitted

Snik AF, Mylanus EA, Proops DW, Wolfaardt JF, Hodgetts WE, Somers T, Niparko JK, Wazen JJ, Sterkers O, Cremers CW, Tjellström A.  Consensus statements on the BAHA system: where do we stand at present? Ann Otol Rhinol Laryngol Suppl. 2005 ;195:2-12.

Snik A, Leijendeckers J, Hol M, Mylanus E, Cremers C. The bone-anchored hearing aid for children: recent developments. Int J Audiol. 2008;47(9):554-9.

Snik, A.F.M. Implantable hearing devices for conductive and sensorineural hearing impairment In: F.-G. Zeng et al. (eds.), Auditory Prostheses: New Horizons, Springer Handbook of Auditory Research 39,  Springer Science+Business Media, 2011:85-108

Snik AFM. Are today’s implantable devices better than conventional solutions for patients with conductive and mixed hearing loss? ENT and Audiology news, 2014;23 (1):105-108

Snik AFM, Vermeulen AM, Brokx JP, Beijk C, van den Broek P. Speech perception performance of children with a cochlear implant compared to that of children with conventional hearing aids. I. The “equivalent hearing loss” concept. Acta Otolaryngol. 1997 Sep;117(5):750-4.

Snik A, Implantable hearing devices for conductive and sensorineural hearing impairment. In: F-G Zeng et al. (eds.) Auditory Prosthesis. New Horizons. Springer Handbook of Auditory Research 39, 2011:85-108

Snik AF, Mylanus EA, Cremers CW. Bone-anchored hearing aids in patients with sensorineural hearing loss and persistent otitis externa. Clin Otolaryngol Allied Sci. 1995;20:31-35

Snik AFM. Hearing impairment in the high-tech era. Inaugural lecture, Radboud University Nijmegen. Thieme MediaCenter, Nijmegen, 2007; ISBN 90-9021439-9.

Snik AFM, Agterberg M, Bosman A. How to quantify binaural hearing in patients with unilateral hearin g using hearing implants. Audiol Neurootol. 2015;20 Suppl 1:44-47

Snik A, Maier H, Hodgetts B, Kompis M, Mertens G, van de Heyning P, Lenarz T, Bosman A. Efficacy of Auditory Implants for Patients With Conductive and Mixed Hearing Loss Depends on Implant Center. Otol Neurotol. 2019 Apr;40(4):430-435.

Sprinzl GM, Wolf-Magele A. The Bonebridge Bone Conduction Hearing Implant: indication criteria, surgery and a systematic review of the literature. Clin Otolaryngol. 2016;41:131-43.

Sprinzl G, Lenarz T, Ernst A, Hagen R, Wolf-Magele A, Mojallal H, Todt I, Mlynski R, Wolframm MD. First European multicenter results with a new transcutaneous bone conduction hearing implant system: short-term safety and efficacy. Otol Neurotol. 2013;34(6):1076-83.

Sprinzl GM, Schoerg P, Muck S, Jesenko M, Speiser S, Ploder M, Edlinger SH, Magele A. Long-Term Stability and Safety of the Soundbridge Coupled to the Round Window. Laryngoscope. 2021 May;131(5):E1434-E1442.

Stelmachowicz, Patricia G.; Hoover, Brenda; Lewis, Dawna E.; Brennan, Marc Is functional gain really functional? The Hearing Journal, 2002;55(11):38-42.

Stenfelt S, Håkansson B, Jönsson R, Granström G. A bone-anchored hearing aid for patients with pure sensorineural hearing impairment: a pilot study. Scand Audiol. 2000;29:175-185

Stenfelt S, Håkansson B. Air versus bone conduction: an equal loudness investigation. Hear Res. 2002;167(1-2):1-12.

Stenfelt S. Bilateral fitting of BAHAs and BAHA fitted in unilateral deaf persons: acoustical aspects. Int J Audiol 2005;44:178-189

Stenfelt S. Transcranial attenuation of bone-conducted sound when stimulation is at the mastoid and at the bone conduction hearing aid position. Otol Neurotol. 2012;33:105-14.

Stenfelt S, Zeitooni M. Loudness functions with air and bone conduction stimulation in normal-hearing subjects using a categorical loudness scaling procedure. Hear Res. 2013;301:85-92.

Stone MA, Moore BCJ, Meisenbacher K, Derleth RP. Tolerable hearing aid delays. V. Estimation of limits for open canal fittings. Ear Hear. 2008;29:601-617

Streitberger C, Perotti M, Beltrame MA, Giarbini N. Vibrant Soundbridge for hearing restoration after chronic ear surgery. Rev Laryngol Otol Rhinol (Bord). 2009;130:83-8.

Strijbos RM, Straatman LV, Calon TGA, Johansson ML, de Bruijn AJG, van den Berge H, Wagenaar M, Eichhorn E, Janssen M, Jonhede S, van Tongeren J, Holmberg M, Stokroos R. Long-Term Outcomes of the Minimally Invasive Ponto Surgery vs. Linear Incision Technique with Soft Tissue Preservation for Installation of Percutaneous Bone Conduction Devices. Front Neurol. 2021 Feb 24;12:632

Sylvester DC, Gardner R, Reilly PG, et al. Audiologic and surgical outcomes of a novel, non-percutaneous, bone conducting hearing implant. Otol Neurotol 2013;34:922-6

Sziklai I, Batta TJ, Karosi T. Otosclerosis: an organ-specific inflammatory disease with sensorineural hearing loss. Eur Arch Otorhinolaryngol. 2009;266:1711-8.

Teunissen EM, Caspers CJI, Vijverberg MA, Pennings RJE, Mylanus EAM, Hol MKS. Long-Term Outcomes of a Percutaneous Wide-Diameter Bone-Anchored Hearing Implant: A Clinical Evaluation of More than 800 Implants. Otol Neurotol. 2024 Jun 1;45(5):e435-e442.

Teunissen EM, Kok HJW, Janssen AM, Hol MKS, Bosman AJ. Evaluation of a super powerful bone-anchored hearing system and its users: A retrospective study. Clin Otolaryngol. 2024 Sep;49(5):670-676.

Thomeer HG, Kunst HP, Cremers CW. Isolated congenital stapes ankylosis: surgical results in a consecutive series of 39 ears. Ann Otol Rhinol Laryngol. 2010;119:761-6.

Thomeer HG, Kunst HP, Cremers CW. Congenital stapes ankylosis associated with another ossicular chain anomaly: surgical results in 30 ears. Arch Otolaryngol Head Neck Surg. 2011;137:935-41.

Thomeer HG, Kunst HP, Cremers CW. Congenital ossicular chain anomalies associated with a mobile stapes footplate: surgical results for 23 ears. Ann Otol Rhinol Laryngol. 2012;121:275-81.

Thomeer HGXM.  Congenital minor ear anomalies. Thesis, Radboud University, Nijmegen, The Netherlands. 2012.

Topsakal V(1), Fransen E, Schmerber S, Declau F, Yung M, Gordts F, Van Camp G, Van de Heyning P. Audiometric analyses confirm a cochlear component, disproportional to age, in stapedial otosclerosis. Otol Neurotol. 2006;27(6):781-7.

Tysome J. et al., The Auditory Rehabilitation Outcomes Network: an international initiative to develop core sets of patient-centered outcome measures to assess interventions for hearing loss. Clin Otolaryngol. 2015;40:512-515

van Barneveld DCPBM, Kok HJW, Noten JFP, Bosman AJ, Snik AFM. Determining fitting ranges of various bone conduction hearing aids. Clin Otolaryngol. 2018 Feb;43(1):68-75.

Van den Bogaert T, Klasen TJ, Moonen M, Van Deun L, Wouters J. Horizontal localization with bilateral hearing aids: without is better than with. J Acoust Soc Am 2006;119:515-26.

van Loon MC, Merkus P, Smit CF, Smits C, Witte BI, Hensen EF. Stapedotomy in cochlear implant candidates with far advanced otosclerosis: a systematic review. Otol Neurotol. 2014;35:1707-14.

Venail F, Lavieille JP, Meller R, Deveze A, Tardivet L, Magnan J. New perspectives for middle ear implants: first results in otosclerosis with mixed hearing loss. Laryngoscope. 2007r;117(3):552-5..

Verhaegen VJ, Mulder JJ, Mylanus EA, Cremers CW, Snik AF. Profound mixed hearing loss: bone-anchored hearing aid system or cochlear implant? Ann Otol Rhinol Laryngol. 2009;118(10):693-7.

Verhaegen VJ, Mylanus EA, Cremers CW, Snik AF. Audiological application criteria for implantable hearing aid devices: a clinical experience at the Nijmegen ORL clinic. Laryngoscope. 2008 Sep;118(9):1645-9.

Verhaert N, Mojallal H, Schwab B. Indications and outcome of subtotal petrosectomy for active middle ear implants. Eur Arch Otorhinolaryngol. 2013 Mar;270(4):1243-8.

Verhagen CV, Hol MK, Coppens-Schellekens W, Snik AF, Cremers CW. The Baha Softband. A new treatment for young children with bilateral congenital aural atresia. Int J Pediatr Otorhinolaryngol. 2008;72:1455-9.

Verhaert N, Desloovere C, Wouters J. Acoustic hearing implants for mixed hearing loss: a systematic review. Otol Neurotol. 2013;34:1201-9.

Verhagen CV, Hol MK, Coppens-Schellekens W, Snik AF, Cremers CW. The Baha Soft Band. A new treatment for young children with bilateral congenital aural atresia. Int J Pediatr Otorhinolaryngol. 2008;72(10):1455-9.

Verheij E, Bezdjian A, Grolman W, Thomeer HG. A Systematic Review on Complications of Tissue Preservation Surgical Techniques in Percutaneous Bone Conduction Hearing Devices. Otol Neurotol. 2016;37:829-37

Vickers D, Briggs J, Lamping W, Andrew R, Bingham M, Toner J, Cooper S, Spielman P, Ghulam H, Nunn T, et al. Medical Safety and Device Reliability of Active Transcutaneous Middle Ear and Bone Conducting Implants: A Long-Term Multi-Centre Observational Study. Applied Sciences, 2023;13(14):8279. 

Vickers et al., Evaluating the effectiveness and reliability of the Vibrant Soundbridge and Bonebridge auditory implants in clinical practice: Study design and methods for a multi-centre longitudinal observational study. longitudinal observational study. Contemp Clin Trials Commun. 2018 Mar 15;10:137-140.

Vogt K, Wasmann JW, Van Opstal AJ, Snik AFM, Agterberg MJH. Contribution of spectral pinna cues for sound localization in children with congenital unilateral conductive hearing loss after hearing rehabilitation. Hear Res. 2020;385:107847.

Vogt K, Frenzel H, Ausili SA, Hollfelder D, Wollenberg B, Snik AFM, Agterberg MJH. Improved directional hearing of children with congenital unilateral conductive hearing loss implanted with an active bone-conduction implant or an active middle ear implant. Hear Res. 2018;370:238-247.

Vogt K, Desmet J, Janssen A, Agterberg M, Snik A. Unexplained variance in benefit of treatment of congenital unilateral aural atresia. Audiol Neurotol 2021; in press

Voss A, Brockmeyer A, Valente M, Pumford J, Wick CC, Durakovic N. Differences in Force Levels, Word Recognition in Quiet, Sentence Reception Threshold in Noise, and Subjective Outcomes for a Bone-Anchored Hearing Device Programmed Using Manufacturer First-Fit, Aided Sound-Field Thresholds and Programmed to DSL-BCD Using a Skull Simulator. J Am Acad Audiol. 2021;32(7):395-404.

Vyskocil E, Riss D, Honeder C, Arnoldner C, Hamzavi JS, Baumgartner WD, Flak S, Gstoettner W. Vibroplasty in mixed and conductive hearing loss: comparison of different coupling methods. Laryngoscope. 2014 Jun;124(6):1436-43.

Wardenga N, Diedrich V, Waldmann B, Lenarz T, Maier H. Hearing Aid Treatment in Patients with Mixed Hearing Loss. Part I: Expected Benefit and Limitations after Stapes Surgery. Audiol Neurootol. 2020;25(3):125-132.

Wazen JJ, Wycherly B, Daugherty J. Complications of bone-anchored hearing devices. Adv Otorhinolaryngol. 2011;71:63-72.

Wazen JJ, Young DL, Farrugia MC, Chandrasekhar SS, Ghossaini SN, Borik J, Soneru C, Spitzer JB. Successes and complications of the Baha system. Otol Neurotol. 2008 Dec;29(8):1115-9.

WHO 2021. World report on hearing. Geneva: World Health Organization; 2021. Licence: CC BY-NC-SA 3.0 IGO.

Wolf-Magele A, Koci V, Schnabl J, Zorowka P, Riechelmann H, Sprinzl G. Bilateral use of active middle ear implants: speech discrimination results in noise. Eur Arch Otorhinolaryngol. 2016;273:2067-207

Yu JK, Tsang WS, Wong TK, Tong MC. Outcome of vibrant soundbridge middle ear implant in cantonese-speaking mixed hearing loss adults. Clin Exp Otorhinolaryngol. 2012;5 Suppl 1:S82-8.

Zeitooni M, Mäki-Torkko E, Stenfelt S. Binaural Hearing Ability With Bilateral Bone Conduction Stimulation in Subjects With Normal Hearing: Implications for Bone Conduction Hearing Aids. Ear Hear. 2016;37(6):690-702.

Zernotti ME, Gregorio MF, Sarasty AC. Middle ear implants: functional gain in mixed hearing loss. Braz J Otorhinolaryngol. 2012;78:109-12.

Zernotti ME, Di Gregorio MF, Galeazzi P, Tabernero P. Comparative outcomes of active and passive hearing devices by transcutaneous bone conduction. Acta Otolaryngol. 2016;136(6):556-8.

Zernotti ME, Chiaraviglio MM, Mauricio SB, Tabernero PA, Zernotti M, Di Gregorio MF. Audiological outcomes in patients with congenital aural atresia implanted with transcutaneous active bone conduction hearing implant. Int J Pediatr Otorhinolaryngol. 2019 Apr;119:54-58.

Zhao S, Gong S, Han D, Zhang H, Ma X, Li Y, Chen X, Ren R, Li Y.  Round window application of an active middle ear implant (AMEI) system in congenital oval window atresia. Acta Otolaryngol. 2016;136(1):23-33.

Zwartenkot JW, Snik AF, Mylanus EA, Mulder JJ. Amplification options for patients with mixed hearing loss. Otol Neurotol. 2014 Feb;35(2):221-6.

Zwartenkot JW, Hashemi J, Cremers CW, Mulder JJ, Snik AF. Active middle ear implantation for patients with sensorineural hearing loss and external otitis: long-term outcome in patient satisfaction. Otol Neurotol. 2013 Jul;34(5):855-61.

Chapter 6. Bilateral application should always be considered

6.1 Introduction

Binaural hearing refers to hearing with two ears. When listening with two ears instead of one, at least three advantages can be distinguished: 1) loudness summation, 2)) directional hearing and 3) binaural squelch. In normal hearing persons binaural hearing is obvious, based on accurate processing of the unilateral inputs, leading to a ‘fused’ percept (binaural hearing). That is not evident for patients using hearing devices. Bilateral input with hearing implants does not automatically result in binaural hearing.

An introduction of the three advantages of binaural hearing:

  1. Loudness summation refers to improved hearing owing to summation of sound as heard by each ear: the input perceived by either cochlea is summed, leading to an increase in loudness of approx. 5 dB.
  2. Localization of sounds, in the horizontal plane. To identify where a sound is coming from, the two ears have to work together. Interaural differences in perceived sounds are detected, caused by the different positions of the two ears with respect to the sound source. These interaural differences concern interaural loudness differences or ILD; owing to acoustic head shadow and interaural time differences or ITD, caused by a difference in arrival times of the sound at the two ears). The head, as an acoustic barrier, attenuates primarily high frequency sounds (> 2000 Hz). As a consequence, high-frequency sounds are being perceived as louder by the ear nearest to the sound source, creating an ILD. Below 1500 Hz, the head shadow causes little attenuation. Instead, ITDs (or, related, interaural differences in phase) are the relevant cue to localize a sound source. Sounds arrive earlier at the ear nearest to the sound source, creating the ITD. Owing to the limited distance between the two ears, the ITD varies between 0 ms (sounds from the front) to a maximum of 0.7 ms (by a sound presented at the very right or left).
  3. Binaural squelch refers to central de-masking. Assume someone is listening to speech coming from the front and ambient noise also coming from the front. Then the noise might mask the speech. However, if the noise is coming from the left, then the speech sounds in the two ears will (still) be in phase (ITD=0). However, not the ambient noise (ITD= 0.7 ms, the maximum value). The differences in ITDs enables the perceptually separation (de-masking) of the speech and the ambient noise, referred to as binaural squelch.

Bilateral application of BTEs in patients with bilateral sensorineural hearing loss leads to bilateral input but not automatically to binaural hearing. The advantage of a second BTE might be worse than the advantage of listening with two ears (see e.g., Bogaert et al., 2006, studying sound localisation). Nevertheless, e.g., Boymans et al. (2008) showed an obvious benefit of bilateral BTEs in a large group of patients with bilateral sensorineural hearing loss, regarding sound localisation and the effective use of head shadow.

6.2 Binaural hearing with ‘BCDs’

When using the air-conduction route, the two ears can be stimulated independently because they are acoustically well isolated from each other. That is not the case for bone-conduction stimulation. The skull transmits the bone-conducted vibrations quite effectively, thus with little damping. Therefore, with one ‘BCD’, not only the ipsilateral cochlea is stimulated but also the contralateral cochlea. This is referred to as ‘cross-stimulation’ (i.e. cross-hearing). The attenuation in dB of the bone-conducted vibrations from one cochlea to the other varies widely with a median of just 5 dB. The variation between subjects is significant as well as within subjects, between the frequencies (Stenfelt, 2012). In other words, for bone-conduction stimulation, the cochleae are poorly acoustically isolated, however, just sufficiently to enable the detection of interaural differences, although, impaired (Stenfelt, 2005).

Table 6.1 presents data on binaural effects when changing from unilateral to bilateral listening in patients using ‘BCDs’ and, as a reference, normal hearing persons. In the latter case, unilateral hearing refers to hearing with one ear occluded. The first row of the Table presents the mean result (and standard deviations) of the ‘binaural’ outcomes as measured in normal hearing persons (Agterberg et al., 2011) and might be considered as standard values. The other rows present an overview of the outcomes of published Nijmegen studies concerning bilateral ‘BCD’ application in bilateral conductive/mixed hearing loss and unilateral ‘BCD’ application in unilateral conductive hearing loss, thus with a second normal hearing ear. All the patients and the normal hearing persons were evaluated with one and the same protocol. 

Binaural summation and speech in ambient noise were assessed using speech; directional hearing was tested using narrow-band noise with a short duration (1s) presented by one loudspeaker out of seven at a location between + 900 and – 900 azimuth (Agterberg et al., 2011). Concerning the binaural summation score, firstly, speech recognition was measured while the patient was listening with one ear (in unilateral cases with their normal ear; in bilateral cases with their left ear only, thus the ‘BCD’ on the right ear was turned off). To quantify binaural squelch, firstly, speech recognition in noise was measured while the patient was listening unilaterally. Speech was presented in front of the patient and the noise at the side of the only hearing ear. Secondly, the ‘BCD’ near the patient’s other ear was activated (in the normal hearing subjects, one ear had been occluded for the first measurement, which is open again for the second measurement). The presented score is the improvement between these two situations. The asterisks indicate values that differ significantly from 0. The next two columns show directional hearing results for low frequency sounds (relying on detection of ITDs) and for high frequency sounds (relying on the detection of ILDs), respectively. The mean absolute error is shown (0 means perfect localisation). Concerning these results: statistically significance was assessed by comparing the scores obtained in the unilateral listening situation (not shown) and bilateral listening situation. 

Let us consider bilateral ‘BCD’ application in bilateral hearing loss. According to the table, row 2, bilateral ‘BCD’ application does lead to the expected improvement in binaural summation (column 1) of, approximately, 4 dB. This result is in agreement with the literature (Janssen et al., 2012; review). Although not measured, binaural hearing in normal hearing subjects is around 5 dB. Obviously, here, the inevitable ‘cross-stimulation’ during bone-conduction stimulation, plays a minor role.

Table 6.1

N Binaural summation (dB) Head shadow & squelch (dB) Directional hearing, 0.5kHz (degrees) Directional hearing, 3 kHz (degrees)
Controls 10 n.a. 4.6 ± 1.6* 7 ± 7^ 8 ± 10^
Bilat CHL, bilat Baha, acquired 17 4.0 ± 2.1* 2.5 ± 1.8* 26 ± 8^ 25 ± 8^
Unilat CHL, acquired 13 2.2 ± 1.5* 3.0 ± 1.7* 16 ± 10^ 20 ± 12^
Unilat CHL, congenital 10 1.3 ± 2.4 1.1 ± 1.9 30 ± 13 31 ± 18

Note. N.a.: not available, CHL: conductive hearing loss.

* significantly different from 0 (p<0.05); ^significantly improved compared to the unilateral listening condition (p<0.05)

Considering row 3 (‘BCD’ application in acquired unilateral conductive hearing loss) a limited but significant binaural summation score was found, however, not for the unilateral congenital cases (row 4). Column 2 shows the binaural squelch/head shadow data. The bilateral conductive hearing loss patients (row 2) profit significantly, however, profit is approx. half that of the normal hearing persons (2.5 vs. 4.6 dB). For unilateral acquired conductive hearing loss, the outcome is approx. the same, while for the congenital cases, no significant benefit was found. 

Columns 3 and 4 comprise the outcome of the sound localisation tests; the mean absolute difference (in degrees) between the perceived location and the real location of the activated loudspeaker are presented. For normal hearing persons, an error of 7-80   was found, for the bilateral ‘BCD’ users approximately 250  and for the patients with unilateral acquired hearing loss using ‘BCD’, approximately 200. These three groups showed significantly improved scores when changing from unilateral hearing (mean unaided scores not presented but in the range of 46-660) to bilateral hearing. Probably, disturbing ‘cross-stimulation’ is the reason for the difference in score between the normal hearing persons and the bilateral ‘BCD’-users. Again, the unilateral congenital cases showed the worst scores, see next paragraph.

6.3 Unilateral congenital conductive hearing loss; elaborated outcomes

The localisation scores of the patients with congenital unilateral conductive hearing loss with their ‘BCD’ active were largely unchanged, namely 340 and 390 for 0.5 and 3 kHz without the ‘BCD’ and  300 and 310 for 0.5 and 3 kHz with the ‘BCD’. The most likely explanation is that these patients, being unilaterally hearing since birth, have learned to use some specific monaural cues rather effectively. For normal subjects such cues are less relevant because they are inferior, rather marginal binaural cues (as elaborated by Agterberg et al., 2012 and Vogt et al. 2020). Vogt et al. even showed that with a percutaneous ‘BCD’ turned on, these patients remained relying on monaural cues for sound localisation, despite the second input (via the ‘BCD’).

In an attempt to understand the moderate results of the patients with congenital onset of their hearing loss, we decided to study whether or not age at intervention played a role. The hypothesis was that maybe a sensitive period exists; the earlier implantation, the better the binaural results would be. Figure 6.1 shows the individual summed outcome of three binaural tests (the head shadow test and the two directional hearing scores) after expressing the outcomes into z-scores, using the standard deviations of the normal hearing persons. Results of 20 patients who used a ‘BCD’ for at least 3 months were available (Nelissen et al., 2015). Figure 6.1 shows the summed z-scores of these 20 patients as a function of age at implantation. For reference purposes the same procedure was applied to the data of the patients with acquired unilateral conductive hearing loss; a mean summed z-score of just over 10 was found in that group. The Figure shows that a value of 10 is achieved by just one of the patients with congenital unilateral hearing loss. Also negative summed z-scores were found, indicating a worse result with ‘BCD’ than without. A significant effect of age at intervention is not seen, while the spread in results is large.

5 Slide2

Figure 6.1. Binaural advantage based on the use of head shadow and sound localisation. Summed z-scores are presented as a function of age at intervention. Data of the ‘BCD’ users that continued using their device and those who stopped are indicated by different symbols.

Fitting of a (percutaneous) ‘BCD’ in patients with a unilateral congenital conductive hearing loss, to enable binaural hearing, might result in a poor outcome; age at implantation doesn’t seem to play a role (in the age-range studied, from 4 years into adulthood).

Furthermore, Nelissen et al. (2015) reported that with a follow-up > 5 years, 65% of the patients had stopped using their ‘BCD’. An association was found between ‘long-term’ non-use and the ‘short-term’ (one-year post-intervention) summed ‘binaural advantage score’. The non-use was not related to age at intervention. It seems that several of these patients were not able to integrate the new input with that of the normal hearing ear, which is needed for a binaural percept. Thus, not surprisingly, the main reason to stop, as indicated by the patients, was complaints about interfering sounds/ambient noise when using the ‘BCD’ in situations with background sounds. Patients with unilateral acquired conductive hearing loss do benefit from ‘BCD’ application; obviously, they profit from their previously developed binaural abilities.

More recently, a review of the literature was published (Vogt et al., 2021) focussing on post-intervention binaural hearing abilities in children with congenital atresia of the ear canal. Interventions included not only a percutaneous ‘BCD’ but also all other types of implantable ‘BCDs’, MEIs like the VSB as well as surgical atresia repair. Surprisingly, all outcomes were disappointing, irrespective of the type of intervention (concerning binaural summation and/or binaural squelch and/or directional hearing). Vogt et al. discussed factors that might have played a role namely 1. age at treatment, 2. whether the treated ear was a ‘lazy’ ear, which means an ear with a not fully maturated neural network, and 3.  the role of the remaining asymmetry in hearing thresholds, post-intervention, as was seen generally. It was argued that the main problem might have been the remaining asymmetry in hearing after the interventions, which typically varied between 20 dB and 35 dB. This asymmetry might hinder the development of binaural hearing.

Concerning the binaural advantage of a MEI in unilateral conductive hearing loss, it should be noted that in contrast to a ‘BCD’, ‘cross-stimulation’ with a MEI is absent. Two publications were found in which outcomes obtained with the VSB MEI were compared to those with ‘BCDs’ in patients with congenital unilateral conductive hearing loss (Agterberg et al., 2014; Vogt et al. 2018). Agterberg et al. compared results of patients fitted either with a ‘BCD’ (n=4) or with a VSB (n=4). All were experienced device users. An anticipated better result with the ‘cross-stimulation-free’ VSB device in a directional hearing experiment could not be established. Same conclusion was drawn from a comparison between the VSB and an active transcutaneous ‘BCD’, the ‘Bonebridge’ device (Vogt et al., 2018). This suggests that ‘cross-stimulation’ is not a major issue for these patients.

Although unilateral congenital hearing loss in children might lead to developmental delays (in speech language development and performance in school), outcomes of today’s treatment options are grossly disappointing (Vogt et al., 2021). Nevertheless, some children experience improved hearing and are happy with that. No conclusive evidence was found regarding the role of age at treatment, degree of asymmetry in hearing after the treatment nor what intervention is the most effective one.

6.4. Bilateral application of ‘BCDs’ and MEI

Bilateral application in bilateral conductive hearing loss is not always considred (Wang et al. 2025), while it works, see Table 6.1, what is also reported by e.g., Janssen et al. (2012; systematic review) and Zeitooni et al, 2016. The latter study was carried out in normal hearing persons; binaural advantages were compared as obtained with stimuli/speech presented bilaterally over headphones (air-conduction route) and, separately, bilaterally by bone-conduction transducers (bone conduction route). The authors reported profit in both conditions, however the binaural advantage (using speech in noise) was approx. two times better with headphone presentation. This factor of 2 is in fair agreement with the difference as presented in column 2, Table 6.1, when comparing the outcomes of the normal hearing persons and the bilateral ‘BCD’ users.

A further question is whether or not (adult) patients with bilateral congenital conductive or mixed hearing loss do profit from fitting bilateral ‘BCDs’. Bosman et al. (2001) reported that the ‘binaural advantage’ as found in their patients with acquired bilateral conductive hearing loss (summarised in row 2, Table 6.1) and those of 6 congenital cases were grossly the same. This might suggest that early binaural auditory experience is not a prerequisite for effective use of bilateral ‘BCDs’ in bilateral conductive hearing loss. More recently, Den Besten et al. (2020) published data on a group of 7 children with bilateral congenital conductive hearing loss, all experienced users of bilateral percutaneous ‘BCDs’. They reported that all children showed improved sound localisation when using two instead of one ‘BCD’, however, localisation scores were moderate in most of the children. Agterberg et al. (2024) demonstrated accurate sound localization with bilateral MEIs in 4 of the 6 included patients. Den Besten et al. also presented data on none-use. Almost 85% of their group of bilaterally implanted children (n=33) used both percutaneous ‘BCD’ devices all the time.

Although transcutaneous ‘BCDs’ have been applied bilaterally, no systematic prospective evaluative study has been found in literature (early 2021).  As far as we know, studies are also missing on binaural hearing after bilateral application of MEIs, like the VSB, in patients with conductive or mixed hearing loss. Wolf-Magele et al. (2016) studied 10 bilateral VSB users, however 6 of them with sensorineural hearing loss, 4 with mixed loss and Koci et al (2016) reported directional hearing measurements in 10 bilateral VSB users; 8 of them had sensorineural hearing loss. Reported results of these mixed groups of VSB users are favourable, but the outcomes are not necessarily representative for patients with conductive or mixed hearing loss. So, specific studies are needed.

Note. A special remark has to be made concerning binaural summation in patients with bilateral conductive hearing loss using (two) devices instead of one. Improved hearing has been observed owing to binaural summation, in the order of 4 dB, see Table 6.1. This is of major importance; as the amplification of ‘BCDs’ is limited owing to the relatively low MPO, see Chapter 4, Figure 4.4 and related text. The ‘gain ratios’ for groups B and C as presented in that figure will improve significantly with the extra 4 dB of gain. For example, at 2 kHz, the NAL-prescribed value of approx. 0.45 is now reached for both devices of group C while for group B a value of 0.38 is found instead of 0.27. 

However, a problem might occur when fitting two devices. When programming two BTEs in bilateral pure sensorineural hearing loss, either BTE is programmed individually. When switching on both BTEs, overstimulation might occur owing to binaural summation. Generally, to deal with that, the volume and maximum output of the two BTEs is lowered; mostly, the software does this automatically. Evidently, for devices with limited MPO like the ‘BCDs’ and MEIs, such a lowering of gain and output is contra-productive and should be avoided.

In bilateral conductive or mixed hearing loss, the application of bilateral ‘BCD’s leads to binaural hearing, although not optimal, with somewhat better result in acquired cases than congenital cases. Device compliance is high.

So far, the patients in the bilateral ‘BCD’ studies had rather symmetric bone-conduction thresholds. It is not clear what might happen in case of asymmetric bone-conduction thresholds. The theoretical problem is that the better ear might be hindered or even over-stimulated by a ‘BCD’ device fitted at the side of the worse cochlea. No data are available on this issue. When using MEIs, such a problem might not occur.

6.5. Binaural hearing and adaptive sound processing

Today’s ‘BCDs’ and VSBs make use of, amongst others, adaptive sound processing like expansion- and/or compression amplification, noise reduction, feedback reduction and/or adaptive directional microphones. Interaural cues (ILDs and ITDs) might be distorted by independently working digital sound processors. Sound processing times are relatively long (3 to 9 ms) compared to the maximum possible ITD (0.7 ms), furthermore the bilateral devices are not synchronized. All these factors might deteriorate interaural cues, resulting potentially in poor horizontal localization (Bogaert et al., 2006; Beck & Sockalingam, 2010). In principle, these advanced sound processing options are not really necessary when fitting patients with good cochlear function (conductive or mixed hearing loss with mild to moderate SNHLc), because the auditory system can still sort out the important information. On the other hand, some of these processing options might be necessary, to deal with limitation of the devices such as feedback, the unnatural position of the microphone, audible microphone noise and the limited MPO.

It is not evident that patients with conductive hearing loss or mixed hearing loss with a minor to moderate sensorineural hearing loss component do profit from advanced adaptive sound processing. Side effects like deteriorated spatial hearing might occur. Conclusive research is lacking.

References

Agterberg MJH, Straatman L, Bruchhage KL, Jürgens T, Hollfelder D, Leichtle A. The Merits of Bilateral Application of Middle Ear Implants in Patients With Bilateral Conductive and/or Mixed Hearing Loss. Trends Hear. 2024;28:23312165241264466.

Den Besten CA, Vogt K, Bosman AJ, Snik AFM, Hol MKS, Agterberg MJH. The Merits of Bilateral Application of Bone-Conduction Devices in Children With Bilateral Conductive Hearing Loss. Ear Hear. 2020;41(5):1327-1332.

Vogt K, Wasmann JW, Van Opstal AJ, Snik AFM, Agterberg MJH. Contribution of spectral pinna cues for sound localization in children with congenital unilateral conductive hearing loss after hearing rehabilitation. Hear Res. 2020;385:107847.

Vogt K, Frenzel H, Ausili SA, Hollfelder D, Wollenberg B, Snik AFM, Agterberg MJH. Improved directional hearing of children with congenital unilateral conductive hearing loss implanted with an active bone-conduction implant or an active middle ear implant. Hear Res. 2018;370:238-247.

Vogt K, Desmet J, Janssen A, Agterberg M, Snik A. Unexplained variance in benefit of treatment of congenital unilateral aural atresia. Audiol Neurotol 2021; in press.

Wang Y, Zhu J, Liu Y, Wang D, Zhao S. Over three-year outcomes of Bonebridge implantation in children and adolescents with congenital bilateral conductive hearing loss. Auris Nasus Larynx. 2025;52(3):207-215.

Zeitooni M, Mäki-Torkko E, Stenfelt S. Binaural Hearing Ability With Bilateral Bone Conduction Stimulation in Subjects With Normal Hearing: Implications for Bone Conduction Hearing Aids. Ear Hear. 2016;37(6):690-702.

 

Chapter 4. Basic considerations; fitting issues and device choice for the elderly

4.1 Introduction

As concluded in the previous chapter, the dynamic range of hearing with BCDs and MEIs is limited due to the low MPO of these devices. Of course, this affects proper fitting of these devices. Standard procedures cannot be used. An attempt is made to develop a fitting model based on an equitable acceptable partial use of the patient’s dynamic range of hearing. The audibility of normal conversational speech is an important factor in this model. The speech level of running speech at a level of 60 dB SPL varies between 20 and 50 dB HL for frequencies from 500 Hz to 6 kHz and between 15 and 45 dB HL at 250 Hz and 8 kHz (Killion and Mueller, 2010). Based on this, a new goal has been set: at least 35 dB of the patient’s ‘dynamic range of hearing’ must be available with the device (Zwartenkot et al., 2014 and Rheinfeldt et al., 2015). The idea is that 35 dB HL is enough to adequately understand conversational speech while the hearing device makes use of wide-dynamic range compression as a kind of automatic volume control. Wide-dynamic range compression increases the gain for low-level sounds while keeping the high-level sounds at loud but comfortable levels (Reinfeldt et al., 2015).

This new criterion is referred to as the ‘DR > 35 dB rule’.

Table 4.1 shows the average MPO data calculated from the data in Table 2.1. The second column indicates up to what mean bone-conduction threshold (average at 0.5, 1, 2, 4 kHz; the PTA4bc or SNHLc) the different devices can be used to comply the ‘DR > 35 dB rule’. An obvious conclusion is that passive tBCDs can only be used effectively in patients with predominant conductive hearing loss.

Table 4.1 The maximum permissible mean bone-conduction threshold for the different types of devices, assuming that the aided dynamic range of hearing is at least 35 dB

Device MPO* Maximum PTA4bc**
Passive tBCD 51 dB HL 15 dB HL
Active tBCD, Osia 80 dB HL 45 dB HL
Active tBCD, Bonebridge 67 dB HL 30 dB HL
Active tBCD, Sentio mini 70 dB HL 35 dB HL
pBCD with standard processor Ponto/Baha 70 dB HL 35 dB HL
pBCD with super-power

processor Ponto/Baha

84 dB HL 50 dB HL
MEI (VSB) 81 dB HL 45 dB HL

* taken from Table 2.2; mean values; **expressed in multiples of 5 dB

A ‘working range’ of 35 dB could be just enough for proper amplification of speech, in combination with wide-dynamic range compression. Wide dynamic range compression’ is the default choice for patients with pure sensorineural hearing loss using BTEs, who have a limited ‘dynamic range of hearing’ due to physiological conditions. When applying BCDs or MEIs, ‘wide-dynamic range compression’ is also used, although the limited ‘dynamic range of hearing’ is not caused by physiological conditions, but technical restrictions of the devices used. Obviously, the starting point when choosing an amplification option should be to choose a device with a high MPO, especially in children, see also Chapter 5.

Important, it should be noted that the maximum PTA4bc values per device type, as shown in the Table 4.1, correspond to the maximum PTA4bc (or SNHLc) values as obtained from clinical studies, shown in Figure 3.3B. Using the trend-curves indicated in that figure, the maximum SNHLc per device type was assessed, viz., about 45 dB HL for VSB, 15-20 dB HL for the two passive tBCDs and around 50 dB HL for pBCDs with super power processor. This match between Figure 3.3B and Table 4.1 validates the MPO-based ‘maximum PTA4bc-values’ for the applications of the different types of devices, as shown in the Table 4.1.

In short, today’s passive tBCDs can just be used effectively in patients with (sub) normal cochlear function, the active tBCDs in patients with PTA4bc between 30 dB HL and 45 dB HL, depending on the device used, and pBCDs in patients with PTA4bc up to 50 dB HL. Notably, most of these values are less optimistic than those claimed by the manufacturers.

4.2. Implanting elderly patients; the best solutions

Due to aging or a progressive (hereditary) cochlear hearing loss, the SNHLc deteriorates over time. The use of data on the age-related hearing deterioration makes it possible to estimate ‘years of effective use’ of a particular device. However, little is known (reported) on age-related hearing deterioration in patients with mixed hearing loss. One paper was identified; Figure 4.1 shows age-related decline in hearing of patients with mixed hearing loss, studied by Iliadou et al. (2006). The authors reported both air- and bone-conduction thresholds for patients suffering form OTSC7. The figure shows the PTA4bc as a function of age; see the first two rows. Data up to 60 years were presented; the indicated PTAbc values at 70 and 80 years are obtained by (non-linear) extrapolations. The thick red lines indicate whether a particular device can (still) be used effectively. For example, the figure shows that the Baha, with its ‘max PTA4bc’ of 50 dB HL (see Table 4.1), can be used effectively for up to about 80 years. The VSB can be successfully applied up to about 75 years.

Probably, the mildest scenario is that of a patient with a stable conductive hearing loss (assume 20 dB HL) and, additionally, presbycusis. Using the median age-related hearing decline for men (ISO 7029; 2017), Figure 4.2 is obtained. In this case, the effective use decreases to 50 years for the ‘Bonebridge’ and 75 years for VSB and the pBCDs while for patients aged 50+, the passive tBCD with magnetic coupling is an effective option up to age 60 years.

Figure 4.1. Effective device use for a patient with progressive hearing loss owing to otosclerosis (OTSC7). SNHLc stands for the PTA4bc. Sophono is a passive tBCD with magnetic coupling

Figure 4.2. Effective device use for a patient with stable conductive hearing loss component and a sensorineural component that deteriorates due to presbycusis

Long-term effectiveness is an often ignored but important factor in counselling, specifically in older patients, to avoid future surgeries where an active tBCD is replaced with a more powerful type or a MEI.

4.3 A simplified prescription procedure, taking the limited MPO values into account

This section deals unilateral fittings of BCDs and MEIs. Choosing the best device for a patient is challenging. First, a note should be made about bone-conduction thresholds. Note that when considering the application of a MEI, well-masked bone-conduction thresholds are essential to ensure that the cochlea of the ear to be implanted is sensitive enough for a successful application. On the other hand, when applying BCDs, unmasked bone-conduction thresholds of the ear to be treated should be considered. Due to the limited transcranial attenuation of the bone-conducted vibrations, the BCD will stimulate the cochlea with the best sensitivity, which can be either the ipsilateral or contralateral cochlea (Stenfelt, 2012). Therefore, the fitting for a BCD should be based on the unmasked bone-conduction thresholds.

To develop a practice-based prescription procedure, we used published data. The selected studies, used in Chapter 3, were reused. For each of the included studies, the gain was calculated as a function of frequency, referred to as the effective gain or EG. The EG is the difference between the unaided threshold (i.e. the bone-conduction threshold) and the aided threshold. Next, the EG was related to the unaided (bone-conduction) threshold. According to the well-validated NAL fitting rule (for sensorineural hearing loss), the gain ratio (in this case the EG divided by the cochlear (bone-conduction) threshold) should equal approx. 0.45 (with some corrections for the low frequencies; Dillon, 2012).

To see if that ratio has been reached, twenty of the previously selected papers could be included, which provided all the data needed for the calculations. Figure 4.3 shows the ‘gain ratios’, presented by device type and by degree of the cochlear loss (PTA4bc here called SNHLc). To address the discontinuities shown in Figures 3.2 and 3.3, the selected studies were divided into three subgroups, according to the mean SNHLc of the included patients. Group 1 consisted of studies in patients with a mean SNHLc below 25 dB HL, group 2 those with mean SNHLc between 25 and 40 dB HL and group 3 those with a mean SNHLc above 40 dB HL; see Table 4.2.

Table 4.2. Subgroups to address the discontinuities as observed in Figure 3.2

Group 1 Group 2 Group 3
PTA4bc < 25 dB HL 25 to 40 dB HL > 40 dB HL
Figure 3.2 shows: Negative EG Intermediate Positive EG
Hypothetical target for amplification Predominant conductive loss: compensate the AB gap! Intermediate ‘Compensate’ the SNHLc as if it was a pure sensorineural loss

 

1

2

3

Figure 4.3. The ‘gain ratio’ as a function of the frequency. Figure 4.3A shows the data from group 1 (see Table 4.2, note that BAHS stands for Baha and Ponto devices together), Figure 4.3B for group 2 and Figure 4.3C for group 3.

Figure 4.3 clearly shows that all the ‘gain ratios’ are below 0.45, the desired value according to the NAL rule (Dillon, 2012). The figure also shows that the ‘gain ratio’ is the highest at 2 kHz, regardless of device type and subgroup. Moreover, differences between the three groups are significant. The differences between the used hearing devices within the groups, are less pronounced.

Such data can be used to develop practice-based prescription procedure. Dillon and colleagues, adopting the NAL rule for patients with mixed hearing loss, argued that in (adult) patients, the entire AB gap doesn’t have to be compensated (Dillon, 2012, Chapter 10.4); a remaining AB gap of 20-25 dB is acceptable. Using 20 dB for a remaining AG gap, Table 4.3 shows the desired aided thresholds (and the resulting) EG as a function of the bone-conduction thresholds.

Table 4.3. Target values as a function of bone-conduction thresholds for the frequencies 1 kHz to 4 kHz

Bone-conduction threshold (dB HL) 0 10 20 30 40 50 60
Desired aided threshold (dB HL)* 20 20 20 20 22 27 33
As a result, the ‘EG’ (dB) -20 -10 -0 10 20 23 27

*either the ‘acceptable AB gap’ of 20 dB or 0.55 times the bone-conduction threshold; the highest value is listed

This practice-based prescription procedure can be applied regardless of the type of device used. The condition is that a device with sufficiently high MPO is selected, as stated in Table 4.1

Importantly, note that for pBCDs, Hodgetts and Scollie (2017) developed a dedicated prescription rule, based on the DSL fitting procedure. See also Voss et al., (2021) and Gascon et al., (2025).

Chapter 3. Basic considerations; the effect of low MPO on amplification and word recognition

What is the effect of a relatively low MPO? Figure 3.1A (top, right) shows the audiogram of a patient with conductive hearing loss. Cochlear (bone-conduction) thresholds are shown. The loudness-discomfort-levels (LDLs; Dillon and Storey, 1998), are also indicated in Figure 3.1B (top, left).

figure 3.1.1    figure 3.1.2

figure 3.1.3

Figure 3.1. Figure A (top left) shows the audiogram; figure B (top, right) present the cochlear thresholds (bone-conduction thresholds) and the LDLs. Figure C (bottom figure) shows the MPO values (labelled with ‘S’) of the applied (passive transcutaneous) ‘BCD’.

By definition, the patient’s dynamic range of hearing is the difference between the cochlear threshold and LDLs. Apparently, Figure 3.1C shows that the MPO values, indicated by symbol S, split the patient’s dynamic range of hearing into two parts; an audible part (45 dB wide, see vertical red arrows) and a second unused part (40 dB wide). The speech area of the patient’s own voice is schematically indicated, with the yellow part being ‘inaudible’.

Intermezzo. The gain of a device is the difference between aided and unaided thresholds. Using the bone-conduction route, the gain, called the effective gain or EG, is the difference between the (BCD) aided thresholds and the bone-conduction thresholds. If the EG=0 dB, this means that with the BCD the air-bone gap is ‘closed’. A negative EG indicates that despite the amplification, a residual air-bone gap exists. See also Appendix 2. End intermezzo.

If the patient is fitted effectively with the BCD (EG=0 dB), then any signal louder that the MPO of the BCD (symbols ‘S’ in Figure 3.1C) cannot be processed properly causing distortions. The patient can prevent distortions by turning down the volume (thus reducing the EG). For instance, in this case, the chosen reduction in EG might be 20 dB. In this way, the louder part of the patient’s own voice (yellow area), which ranges from 50 to 70 dB HL, is perceived undistorted, now ranging from 30 to 50 dB HL. Consequently, the patient’s own speech fits nicely into the BCD-aided dynamic hearing range. However, that 20 dB attenuation EG of – 20 dB negatively affects speech recognition of someone else, who talks at a normal conversational level. The ‘negative EG’ of 20 dB appears in the audiogram as a ‘remaining AB-gap’ of 20 dB. This example highlights that it is better to use a BCD with a high MPO, even in patients with a (sub)normal sensorineural hearing loss component.

The next question is: what is the volume setting (chosen amplification) in daily practise by BCD users with different degrees of cochlear hearing loss? This question was answered using our Baha database by reviewing the data of 89 patients who used Baha devices (Snik et al., 2004). At that time, Baha devices with (mainly) linear amplification were used (Baha Compact, Classic, HC220, Cordelle) and with a volume wheel that allowed the patients to choose their preferred amplification. (Remark. In the next figures the cochlear hearing loss or the sensorineural hearing loss component is referred to as the SNHLc. The SNHLc stands for the mean bone-conduction threshold at 0.5, 1, 2 and 4 kHz averaged, often referred to as the PTA4bc.)

3.2_nieuw

Figure 3.2 The ‘Baha’-aided thresholds (at 0.5, 1 and 2 kHz averaged) as a function of the SNHLc (or PTA4bc). For clarity, individual data of the 89 Baha users were grouped into 12 classes of each 5 dB wide. The diagonal line represents the bone-conduction thresholds

Figure 3.2 shows that patients with (sub)normal cochlear function (low SNHLc) set the volume of their Baha such that the EG is negative: the ‘Baha’-aided thresholds are worse (higher) than the bone conduction thresholds (shown in the figure as data points above the diagonal). This results in a remaining AB-gap of around 10-15 dB, which is seen for SNHLc values up to approx. 25 dB HL.

Patients with more severe SNHLc need ‘positive’ amplification to be able to properly hear other subjects talking, which, according to Figure 3.2, is found for SNHLc > 30 dB HL.

It should be noted that with increasing SNHLc, patients used the more powerful Baha sound processors. Per patient, the type of Baha device used was documented and, consequently, its MPO was known. These MPO values were used to obtain the ‘input level at saturation’ or the sound level at which the MPO was reached. Louder levels lead to distortions. This ‘input level’ is simply the MPO value minus the amplification provided by the Baha (output = input + amplification). The amplification itself is easily obtained by subtracting the Baha-aided thresholds from the bone-conduction thresholds. These calculated (individual) input levels at which the Baha devices saturated, were (again) grouped in the 12 classes, presented in Figure 3.2 as the blue line. The figure clearly shows that the input level at saturation restricts the dynamic range of hearing with the Baha significantly. Before discussing the consequences, Figure 3.2 was redrawn, see Figure 3.2bis.

 3.2_bis  3.2_bis_2

Figure 3.2bis Redrawn of Figure 3.2 with more details, see text

The left graph of Figure 3.2bis shows the residual capacity of ear of patients with SNHLc between 0 and 60 dB HL. As mentioned earlier, the diagonal line presents the cochlear thresholds while the dotted line presents the patient’s LDL. The graph on the right shows the aided situation after Baha fitting, taken from Figure 3.2. The difference between aided thresholds and input level at saturation is the patient’s aided dynamic range of hearing or their ‘auditory window’ to the world.

The Baha-aided dynamic range of hearing is limited; for example, with a SNHLc of 45 dB HL, the patient’s unaided dynamic range of hearing is 50 dB HL (95-45 dB HL, left graph, red arrow) and with the Baha, the dynamic range is only 35 dB HL (40 to 75 dB HL, right graph, red arrow).

The subfigure on the right side schematically shows sounds from the real auditory world. Let us consider normal conversational speech. That level fluctuates between 25 dB HL and 55 dB HL (indicated in the subfigure, right side; according to Killion & Christensen, 1998 based on speech with an overall level of 65 dB SPL). By combining the right subfigure and the graph on the right shows that in the current patient group, the 30 dB speech range is only fully audible when the SNHLc is less than 15 dB HL, while at a SNHLc of 60 dB HL, the audible proportion of the speech range has decreased to just 5 dB HL.

In addition, we learn from Figure 3.2 that the patients, regardless of their SNHLc, adjust the volume of their device so that the device saturates at input sound levels of 70-75 dB HL. Indeed, this choice allows for undistorted perception of even the loudest parts of normal conversational speech. Thus, the figure suggests that the patients did not compromise on the input level at saturation, but did make compromises with respect to the softest speech sounds they could hear.

The basic rule for the desired gain in pure sensorineural hearing loss equals approximately half the sensory neural hearing loss (e.g., Dillon, 2012). As the figure shows, that has not been achieved at all. For example, at a SNHLc of 45 dB HL, the EG is only 7 dB (see Figure 3.2). It is concluded that the limited MPO plays an important role in the low gain setting chosen by the patients.

Until now, the analysed data concerned Baha devices. Figure 3.3 shows similar data from all implantable systems available (systematic literature search performed mid-2017), including besides the Baha the percutaneous Ponto device (Oticon Medical, Askim, Sweden), an active tBCD, viz. the Bonebridge device (Med-El, Innsbruck, Austria), two passive tBCDs; the Baha Attract (Cochlear BAS, Gothenburg, Sweden) and Sophono (Medtronic, Boulder, Co, USA) and the Codacs device (Cochlear, Mechelen, Belgium); the latter two devices are no longer on the market (details on the literature search are presented at the end of this section; see Note, just before section 3.2).

Altogether, 48 studies were included in this analysis. Figure 3.3A shows the mean aided free-field threshold at 0.5, 1, 2 and 4 kHz as a function of the mean bone-conduction threshold, (the SNHLc or PTA4bc). This figure again shows that, in general, when the SNHLc is below 25-30 dB HL, a negative EG, thus a remaining AB gap is found (since the data points are above the diagonal). This observation corresponds to Figure 3.2.

Figure 3.3B presents the same data as Figure 3.3A. This time, trend-curves are presented, drawn by eye through the data points for each type of device. Given the blue line, representing the trend for VSB, the aided thresholds are at a fairly stable level (25-30 dB HL) up to a SNHLc of approx. 45 dB HL. Above that value, the aided thresholds appear to deteriorate exponentially, suggesting that the device is no longer powerful enough for such degrees of the SNHLc. For the passive tBCDs (Baha Attract / Sophono), similar trends can be seen, however, with lower cut-off SNHLc value, between 15-20 dB HL. Due to their MPO values, evidently, these passive tBCDs and the VSB are not equivalents. No clear limit is found for the pBCDs (Baha/Ponto devices). This is due to the availability of three types of processors (standard, power and superpower) that were used, depending on the patients SNHLc. However, the figure suggests that the cut-off value for pBCDs lies around 50 dB HL. For the Bonebridge, no clear cut-off value was found up to 25 dB HL. Ignoring the trends, Figure 3.3A shows a rather large variability (vertical distribution) between studies. This may be clinic-related, for example due to differences in fitting protocols.

Slide3

Figure 3.3A. Mean aided free-field threshold (0.5, 1, 2, 4 kHz) as a function of the mean SNHLc (mean bone conduction thresholds at the same frequencies) as obtained from 48 studies using the percutaneous Baha/Ponto devices (black diamonds), the VSB (big blue dots), the ‘Bonebridge’ (red squares), the Baha Attract (green triangles), the Sophono (small black dots) or the Codacs (green circles). The Baha Attract and Sophono are passive tBCDs

Slide2

Figure 3.3B. The same figure as Figure 3.3A, however, this time trend lines are indicated for the VSB and the passive tBCDs

It might be argued that updates/upgrades of the devices might have led to better aided thresholds and better EG. However, an overview of more recent data (2018-2025), shows similar outcomes, see next section (3.2).

Note. To find relevant studies, it was decided to first use the studies that had been previously selected by authors who published systematic reviews. This concerned reviews by Colquitt et al., 2011; Verhaert et al., 2013; Nadaraja et al., 2013; Ernst et al., 2016; Sprinzl and Wolf-Mangele, 2016 and Dimitriades et al., 2016. Of all the included studies, only those that presented air- and bone-conduction thresholds and aided free-field thresholds were selected. Duplicates were removed as well as studies with five patients or fewer. Unpublished presentations were also excluded. Due to low numbers, additional searches in the literature were carried out. The search period was 2006-2017. Concerning the percutaneous Baha and Ponto devices, using the reviews, nine publications met the inclusion criteria (Kompis et al., 2007; Priwin et al., 2007; Flynn et al., 2009; Fuchsman et al., 2010; Ricci et al., 2011; Marsella et al., 2011; Bouhabel et al., 2012). Concerning the VSB, using the reviews by Verhaert et al. and Ernst et al., 14 publications were included (Beltram et al. 2009; Streitberger et al. 2009; Colletti et al. 2011; Baumgartner et al., 2010; Frenzel et al., 2010; Beleites et al. 2011; Bernardeschi et al., 2011; 2013). Sprinzl and Wolf-Mangele (2016) reviewed the outcomes of ‘Bonebridge’ application. Only two studies met the criteria (Sprinzl et al., 2013; Ihler et al., 2016). Dimitriades et al. (2016) reviewed the transcutaneous Baha Attract; three studies were included (Iseri et al., 2014; 2015; Briggs et al., 2015). Since the number of included publications regarding the Baha Attract and the ‘Bonebridge’ were low and the publications with the percutaneous Baha/Ponto outcomes were rather dated (published before 2012), we performed an additional literature search, using Pubmed. For Bonebridge and Baha Attract, search terms were the respective device names together with ‘2016’; concerning ‘Bonebridge’, four studies were added (Gerdes et al., 2016; Baumgartner et al., 2016; Smerber et al., 2016; Zernotti et al., 2016), for Baha Attract, no further study with relevant audiometric data was found. Concerning Baha/Ponto, the sound-processor’s names were used; the search period was 2011-2017. In contrast to the other devices, number of included patients had to be ten or more. Seven studies were identified (Pfiffner et al., 2011; Flynn et al., 2012; Bosman et al., 2013; Kurz et al., 2013; Desmet et al., 2013; Kompis et al. 2014; Gerdes et al. 2016). Concerning the Sophono device, a systematic review has not yet been published. Therefore, we searched for studies using only ‘Sophono’ as search term and applying the introduced inclusion criteria. Six studies were included (Siegert and Kanderke, 2013; Sylvester et al., 2013; Hol et al., 2013; Denoyelle et al., 2015; Shin et al., 2016; Magliulo et al., 2015). Finally, data obtained with the Codacs device were included, which is the most powerful implantable device (Zwartenkot et al., 2014). Searching PubMed, three papers were identified comprising audiological data obtained in more than five patients (Busch et al., 2013; Lenarz et al., 2013; 2014).

3.1 Update 2026, The effect of low MPO on aided thresholds and word recognition, a further elaboration of the data presented in section 2.3.

Based on the MPO values in Table 2.2, the audiological capacity of the BCDs and MEIs can be ranked. While MPO offers a theoretical limit, real-world performance depends on aided thresholds and, relatedly, speech recognition. To verify that the differences in MPO translate into functional benefit, a targeted literature review was conducted. Contrary to the design used in the previous section, it was decided to just select studies on patients with conductive hearing loss, to create homogeneous groups of BCD and MEI users.

The aim of the review was to collect aided hearing thresholds in patients using a pBCD or active tBCD, unilaterally. The search period was 2018-Februari 2025. Studies should present bone-conduction thresholds and aided free-field thresholds.  For more details on the selection process, see Snik et al. (2026). Recently, more data have been gathered, following the same criteria, on passive tBCDs and the VSB.

Table 3.1 shows an overview of the number of enrolled studies and patients by device type, as well as the weighted average PTA4bc and ‘aided PTA4’ (averaged over studies, taking into account the number of patients). The calculated mean EG4 (0.5 to 4 kHz) is also displayed. A similar result is found when ranking the mean aided PTA4 and comparing it with that of the MPO (Table 2.2).

Table 3.1. The included number of studies and patients as well as the (weighted) mean aided PTA4 (adjusted for the number of patients per study) as well as the mean EG

Device N

Study

N

(pats)

Mean PTA4bc

dB HL

Mean aided PTA4

± s.d.*; dB HL

Mean EG4

dB

Passive tBCD 12 301 11 29±3 -18
Bonebridge 10 296 9 28±4 -19
Osia 4 239 14 25±2 -11
VSB 4 81 10 27±2 -17
pBCD 8 224 12 20±2 -8

*spread in outcomes; best estimates

In accordance with the previous section (3.1), negative EG values are found again in these patients with subnormal cochlear function. Clear differences can be seen between devices, with the aided PTA4 varying between 20 dB HL and 30 dB HL. This has consequences for speech recognition. The model presented by Killion and Mueller (2011; based on the speech intelligibility index) makes it possible to assess the word recognition score (WRS) based on the aided thresholds. For speech at a general level of 60 dB SPL (normal conversational level) with the aided thresholds at 20 dB HL, the WRS is 95% and for 30 dB HL, 85%. For slightly softer speech, although still within the range of normal conversational levels (50 dB SPL), the WRS is 90% and 70%, respectively.

In summary, there are clear differences in the audiological capacity between the different types of BCDs and VSB, not only in technical terms (MPO) but also in terms of speech recognition. Although not further explained here, device noise also pays a role. In contrast to the active tBCDs and the VSB, device noise from pBCDs is not audible (e.g., Carlsson & Hakansson, 1997).

Apparently, the pBCD has a unique audiological capacity. However, there may be practical, medical, surgical, financial and/or esthetical reasons for choosing one of the other amplification options. See next chapter. 

Chapter 2. Basic considerations; Amplification options for conductive and mixed hearing loss

2.1 Introduction

Today, for patients with conductive hearing loss or mixed hearing loss, who need amplification, common options include: 1. conventional air-conduction devices, for example behind-the-ear devices (BTE), 2. bone-conduction devices (BCD) and 3. active middle ear implants with their actuator coupled to one of the cochlear windows. These three options are neither equivalent nor interchangeable in terms of medical or anatomical limitations and their technical capacities. This chapter discusses the audiological capacities of all these devices.

2.2. Air-conduction versus bone-conduction hearing aids; capacity of these devices

Basically, devices that use either the air-conduction or bone-conduction route can be considered as amplification options for patients with conductive or mixed hearing loss. In general, for patients with pure sensorineural hearing loss, BTEs are the first choice, fitted according to standardized procedures (e.g., Dillon, 2012).

Figure 2.1. A conventional BTE (behind-the-ear) hearing aid with earmould. Source: Internet

However, when an AB-gap is present (gap between the air- and bone conduction hearing thresholds), a significant portion of the BTE output (equal to the width of the air-bone gap) is ‘lost’ before the amplified signal reaches the cochlea. As a result, the remaining output, to ‘compensate’ for any cochlear hearing loss, may become too low or even absent. Take, for example, a patient with mixed hearing loss with an AB-gap of 50 dB, who uses a standard BTE. According to the documentation, the MPO (maximum power output) of that BTE is 114 dB SPL and the maximum gain is 48 dB. Consequently, the MPO as perceived by the patient is 114-50, i.e. 64 dB SPL, and the maximum gain is (48-50), so -2 dB. Note that the 64 dB SPL is equal to the sound pressure level of normal speech. The relatively louder speech elements cannot be amplified properly, leading to distortions. Therefore, when applying BTEs in mixed hearing loss, only the most powerful BTEs should be considered and if the AB-gap is large, the benefit might be limited. Wardenga et al. (2020), who studied patients with otosclerosis or tympanosclerosis (no contraindication to the use of a BTE), showed that the width of the AB-gap is a key success factor for a BTE fitting.

The other amplification option is a hearing device that uses the bone-conduction route; a bone-conduction device, or BCD. A BCD consists of a microphone, amplifier and actuator (vibrator), usually in one housing. The BCD is placed behind the ear, for example secured with a head band; see Figure 2.2. Such a BCDs is called a passive transcutaneous BCD (passive tBCD) because the skull is vibrated though the (intact) skin. Alternatively, the BCD can be held in place with an adhesive or a magnetic coupling. In the latter case, a coupling magnet must be implanted under the skin (Table 2.1 presents an overview of passive tBCDs on the market, end 2025).

From an efficiency perspective, the bone-conduction route for sounds is less effective than the air-conduction route. Amongst others, Skoda-Türk and Welleschik (1981) showed that the air-conduction route is even 50 dB more effective than the bone-conduction route. In other words, the first 50 dB produced by the amplifier of a passive tBCD are ‘lost’, making this solution as (in)effective as the BTE for a patient with a 50 dB AB-gap.

Figure 2.2. Passive tBCD; the Baha Softband. The headband keeps the device in place. Source: Cochlear company, with permission

A new, more effective ’BCD’ was developed in the mid 1980s (Hakansson et al., 1984); the (adapted) BCD is connected to the skull via a skin-penetrating titanium implant, anchored in the skull bone (see Figure 2.3). As shown by Hakansson et al. (1984) this ‘percutaneous’ bone-conduction pathway is approximately 15-20 dB more efficient than the transcutaneous pathway, due to the absence of the vibrations-attenuating skin and subcutaneous layers. This device is called a percutaneous BCD or pBCD.

By combining the findings by Hakansson et al. and those of Skoda-Türk and Welleschik suggests that the percutaneous route is only 30-35 dB (the reported 50 dB minus 15-20 dB) less effective than the air-conduction route. This implies that if a hearing-impaired subject has an AB-gap of 30-35 dB, a BTE or pBCD might be equally effective with the pBCD as the most effective solution for patients with an AB-gap of more than 35 dB and the BTE if the AB-gap is below 30 dB. Indeed, de Wolf et al. (2011) showed that this theoretical ‘cross-over’ point is realistic.

Figure 2.3. The Bone-anchored hearing aid or Baha with its actuator connected to the skull via a titanium percutaneous implant. Source: Cochlear Company, with permission 

Table 2.1. Available BCDs, early 2026

Passive tBCDs Coupling Manufacturer
Baha Softband Headband Cochlear BAS, Goteborg, Sweden
Ponto Softband Headband Oticon Medical, Askim, Sweden
Adhear Adhesive Med-El, Innsbruck, Austria
Baha Attract Magnetic Cochlear BAS, Goteborg, Sweden
Percutaneous BCDs
Baha percutaneous Cochlear BAS, Goteborg, Sweden
Ponto percutaneous Oticon Medical, Askim, Sweden
Active tBCDs
Bonebridge Med-El, Innsbruck, Austria
Osia Cochlear BAS, Goteborg, Sweden
Sentio mini Oticon Medical, Askim, Sweden
MEI
Vibrant Soundbridge Med-El, Innsbruck, Austria

2.3 The maximum output-level of bone-conduction hearing implants and other implantable amplification options

BTEs cannot be used by hearing-impaired patients in case of atresia of the ear canal or stenosis or chronically infected ears (see Chapter 5). Then, BCDs are the next option. Carlsson and Hakansson (1997) studied the gain and MPO of the first pBCD on the market, the Baha device. The MPO is the loudest sound that a device can produced. They studied the standard Baha (HC200 at that time) and showed that the MPO was 100, 112, 102, 95 dB FL (decibel-force-level) at 0.5, 1, 2 and 4 kHz, respectively, as measured on a so-called ‘skull simulator’. These data can be expressed in dB HL, using the so-called RETFLdbc (Reference Ear to Force Level for direct bone conduction; Carlsson and Hakansson, 1997), resulting in MPOs expressed in dB HL. In this case (HC200), the MPO was 53, 66, 78, 69 dB HL on the four frequencies, respectively, with an average of 67 dB HL. For reference, the average MPO of a standard BTE can exceed 110-120 dB HL. Carlson and Hakansson also studied the noise floor of that Baha, which was below 0 dB HL, so inaudible.

The MPO measurements have been reproduced and extended using other/newer types of pBCDs by Zwartenkot et al. (2014). They also studied the Ponto pBCD, a competitor of Baha (see Table 2.1). There are three types of pBCDs available: the standard, power, and superpower pBCD. The MPO of the standard and superpower pBCD are presented in Table 2.2; as demonstrated by e.g. Ellsperman et al. (2021), the Baha and Ponto pBCDs are equally effective, therefore, in Table 2.2, the reported MPOs of Ponto and Baha devices are presented combined.

The method developed by Zwartenkot et al. (2014) could also be used to measure the MPO of passive tBCDs. Since the transcutaneous coupling is about 15-20 dB less effective than the percutaneous coupling (Hakansson et al., 1984), it is expected that the MPO of passive tBCDs can be approximately 50 dB HL (the ‘percutaneous’ MPO of 67 dB HL minus 15-20 dB). Indeed, Zwartenkot et al. reported an average MPO of 53 dB HL for the Sophono device, a passive tBCD with magnetic coupling. Here, it should be noted that studies found no systematic differences in audiological outcomes between the available types of passive tBCDs (with headband, adhesive or magnetic coupling); for example, as reported by Gawliczek et al. (2018), who compared the Baha Softband to the adhesive coupling, and Den Besten et al. (2019), who compared the Baha Softband to the magnetic coupling. Therefore, the MPO data from these different types of passive tBCDs are combined in the following. Table 2.2 shows reported MPO values.

2.3.1 New development

Since skin problems around the percutaneous implant were not uncommon, the so-called active tBCDs were developed, consisting of an externally worn sound processor with a separated actuator that is implanted subcutaneously. Actuator (with a receiver) and the processor communicate via an induction link. The main advantage of active tBCDs is that the skin is closed. The first active tBCD on the market was the ‘Bonebridge’ device, see Figure 2.4. The currently available active tBCDs are listed in Table 2.1. Mertens et al. (2014) measured the MPO of the Bonebridge device according to the protocol described by Zwartenkot et al. (2014) and reported a mean MPO value of 63 dB HL. Recently, Ghoncheh et al. (2024) reported a value of 65 dB HL, thus a similar outcome.

Table 2.2. Objective measurement of the MPO of several hearing devices for patients with conductive and mixed hearing loss

Device MPO measured in patients* Reference
Passive tBCD 45-58 dB HL Zwartenkot et al., 2014

Van Barneveld et al., 2018

Active tBCD, Osia 80 dB HL ** Fierens et al., 2022
Active tBCD, Bonebridge 64-67 dB HL Mertens et al., 2014; van Barneveld

et al., 2018; Choncheh et al, 2024

Active tBCD, Sentio mini 70 dB HL**
pBCD with standard processor Ponto/Baha 68-71 dB HL Zwartenkot et al., 2014

Van Barneveld et al., 2018

pBCD with super-power

processor Ponto/Baha

83-85 dB HL Zwartenkot et al., 2014

Van Barneveld et al., 2018

MEI (VSB) 80-82 dB HL Zwartenkot et al., 2014

Maier et al., 2023

* average value at 0.5, 1, 2 and 4kHz; range between studies is presented, **best estimates, see text and note at the end of this chapter

Two more active tBCDs have been introduced more recently; the Osia device and the Sentio device, see Tables 2.1 and 2.2. Note that the MPO of the Osia and Sentio devices have not been reported yet (end 2025), however, using cadaver heads, Fierens et al. (2022) showed that MPO of the Osia is comparable to that of a pBCD with super-power processor; therefore, the estimated MPO value of 80 dB HL is listed in the Table. About the Sentio mini device, the estimated MPO is 70 dB HL. The difference in MPO between the Osia and Sentio mini is due to battery life: the relatively high power-consumption of active tBCDs plays a role; see note at the end of this section 2.3.

Figure 2.4. The ‘Bonebridge’ device; an active transcutaneous ‘BCD’ with its actuator implanted in the mastoid area. The actuator is driven by an externally worn soundprocessor. Source: Med-El, with permission

   

Figure 2.5 The ‘Vibrant Soundbridge’ in classical application (for sensorineural hearing loss; left figure) with its actuator (FMT) coupled to the incus (shown enlarged) and the alternative application for conductive or mixed hearing loss with the FMT coupled to the round window membrane. Source: Med-El.

In 2006, Colletti and his co-workers published an article about the coupling of the actuator of the ‘Vibrant Soundbridge’ middle ear implant (VSB) directly to one of the cochlear windows. This new option is intended for patients with conductive or mixed hearing loss. Figure 2.5 shows the VSB in its classic application with its actuator (called the FMT; floating-mass-transducer) connected to the intact ossicular chain (to the incus), allowing the FMT to move in parallel with the stapes. The right-hand part of Figure 2.5 shows the FMT in its alternative position, coupled to the round window of the cochlea, according to Colletti et al. (2016). First report on the MPO of the adapted VSB for application in conductive and mixed hearing loss showed a MPO between 65 dB HL and 88 dB HL (Zwartenkot et al., 2014). This variability is likely due to the variable effectiveness of the individual coupling of the actuator to the cochlea (Shimizu et al., 2011). As an estimate, ignoring Zwartenkot et al.’s two cases with, most likely, a poor coupling, an average MPO of 82 dB HL was found. Recently, Maier (2023; their most recent cases) reported a mean MPO of 80 dB HL, so quite like that reported by Zwartenkot et al., see Table 2.2.

Based on the MPO values presented in Table 2.2, the capacity of the devices can be ranked.

Evidently, the pBCD has the highest MPO, allowing for the best aided free-field thresholds and thus aided speech scores; see next chapter.

Note.  According to the IEC 60118-9 standard, the MPO of active tBCDs is expressed in dB force level (dB FL or dBmN), referred to as the OVFL.  The average OVFL at 0.5, 1, 2 and 4 kHz of the Osia device is 111 dBmN (Cochlear, 2023), that of the Sentio, 109 dBmN (Oticon, 2024), so quite comparable. A conversion to dB HL is not available yet, however, Fierens et al. (2022) showed in cadaver head, that the MPO of the Osia device was comparable to that of a pBCD with a super-power processor, i.e. about 80 dB HL (see Table 2.2); the MPO of the Sentio will be similar. However, the current consumption of the Osia (and other active tBCDs) is relatively high, for example, the battery life of the Osia device is only 20-30 hours (Cochlear 2023). As for the Sentio device, the currently available soundprocessor is the Sentio mini. To extend battery life, the output was reduced by 12 dB resulting in an the OVFL of 98 dBmN (Oticon, 2024). This increases the battery life to 60-70 hours. Owing to that reduction in output, the MPO of the Sentio mini is approx. 70 dB HL.

Finally, a factor that affects the level of the aided thresholds is the use of expansion (reduced amplification for soft sounds, to make device noise inaudible. To hear the device noise is particularly an issue for patients with conductive hearing loss (thus normal cochlear sensitivity). Expansion suppresses device noise, but it also leads to worse aided thresholds.

 

Chapter 8. Sensorineural hearing loss

8.1 Auditory implants for moderate to severe sensorineural hearing loss

To help patients with sensorineural hearing loss, conventional air-conduction hearing aids are the first choice (e.g. behind-the-ear devices or BTEs; Figure 2.1). If the sensorineural hearing loss is not too severe, it is possible to make conversational speech audible with BTEs. For patient with severe to profound sensorineural hearing loss (> 70-80 dB HL), BTEs might no longer be sufficiently effective due to the impossibility of proper restoration of speech audibility and severe cochlear distortions. Then, cochlear implantation becomes the next treatment option (Hoppe et al., 2015; Huinck et al., 2019), as further explained in the next chapter, section 9.1. An alternative amplification option for patients with hearing loss below 70-80 dB HL is a middle-ear implant (MEI, Verhaegen et al., 2008). High sound fidelity, high output with less feedback issues and the lack of an occluding earmold were the claimed advantages of MEIs over BTEs (e.g. Goode, 1995). Since the late 1990s, different types of MEIs have been introduced for patients with sensorineural hearing loss (Snik, 2011). Nowadays (early 2023), there are two semi-implantable MEIs on the market, namely the Vibrant Soundbridge (VSB; Figure 2.6, chapter 2) and the Maxum device (Ototronix, USA; Pelosi et al., 2014). The latter device comprises a permanent magnet that is surgically connected to the stapes. The driver is an electro-magnetic coil with electronics placed in a deep in-the-ear hearing device. Early 2020, a competitor of the VSB, the semi-implantable Cochlear MET, was withdrawn from the market.

In addition to semi-implantable MEIs, there is one fully implantable MEI is on the market, namely the Esteem device (Envoy Medical, St. Paul, MN, USA; Marzo et al., 2014). Another fully implantable MEI was the Cochlear Carina, which was withdrawn from the market early 2020. Several systematic reviews have been published that focussed on the claimed benefits of MEIs over BTEs. Pulcherio et al. (2014) concluded that, when comparing the Carina and Esteem devices with BTEs, no structural audiological benefit was found. It should be noted that these implantable devices are 1. expensive and 2. include (implant) surgery and 3. revision surgery is needed (every 5-10 years) to replace the battery; therefore, the cost-benefit ratio of these devices is unfavourable compared to BTEs. Another review by Kahue et al. (2014), comparing audiological outcomes obtained with the VSB and Maxum devices on the one hand and BTEs at the other, concluded that an added value of the MEIs could not be established. Since these reviews were published, screening the more recent literature, we found only one study with the Maxum and one with the Esteem device, published by manufacturer-independent research groups. Barbara et al. (2018b) reported on three patients using the Maxum; compared to BTEs, two patients had equal speech recognition scores, and the third patient had a 10% higher word score with the Maxum. Regarding the Esteem device, Barbara et al. (2018a) published an overview of their outcomes. However, these data had been published earlier (and were included in the systematic reviews by Kahue et al. and Pulcherio et al.).

It can be concluded that current MEIs do not have convincing audiological advantages compared to BTEs and are expensive solutions. An exception, however, concerns patients who don’t tolerate earmolds of BTEs due to, for example, chronic external otitis (consensus statements; Magnan et al., 2005). Edfelt et al. (2014) showed that for such patients the application of the VSB is cost effective. The recent systematic reviews, introduced above, indicate that the consensus statement remains valid. Note that according to the set-up of the Maxum, that device is not an option for patients with external otitis. No articles could be found about the Esteem device applied in patients with external otitis.

There is no evidence that MEIs in patients with sensorineural hearing loss have led to better speech recognition than BTEs. However, the VSB can be applied (cost-)effective in patients with sensorineural hearing loss and chronic external otitis.

BCDs have also been used to help patients with sensorineural hearing loss and chronic external otitis. With a BCD, the ear canal remains open. Bone-conduction works but it is not really effective (see Chapter 2). Indeed, studies with the powerful percutaneous BCD (Baha), applied to such patients showed that even at maximum amplification the result was insufficient (Snik et al., 1995). Stenfelt et al. (2000) came to a similar conclusion.

To help patients with sensorineural hearing loss who need amplification, studies showed that BCDs have insufficient capacity.

8.2 Capacity of VSB with today’s sound processors 

In contrast to the Maxum and Envoy Esteem devices, the VSB has been used extensively. With the more recent sound processors (Amade or Samba), improvements in sound quality have been achieved compared to the older processors. In terms of speech recognition, Figure 8.2 shows the percentage correct score (MPS; phoneme scores using monosyllables), at normal conversational level (65 dB SPL). Fourteen patients were tested, using these upgraded VSB processors. Patients with ski-slope audiograms were excluded, after Verhaegen et al., 2008. The lines drawn are from the latter study, based on the speech performance of VSB users with an earlier generation of sound processors. The figure shows that the new data points are close that line, but better. Thus, improved speech recognition, as was also reported by Busch et al. (2016).

The data shown in the figure were obtained in 2015. MPS values were available but also the (mean) word scores (WRS), which was 68% ± 25%. The mean PTA3bc was 48 ± 11 dB HL. More recently, Brkic et al (2022) published similar data from 58 VSB users with a mean PTA3bc of 51 dB HL. They reported a mean WRS which was 7% worse, suggesting that our 2015 data is not outdated yet.

As a reference, Verhaegen et al. showed that with a BTE, applied to patients with similar degrees of (sensorineural) hearing loss, the mean MPS was significantly higher. At a mean sensorineural hearing loss of 65 dB HL, the mean MPS with BTE was 90% versus 70% with VSB (see next chapter, section 9.1.2.)

 

SNHL PS65vsPTA enkel VSB

Figure 8.1. Speech recognition-in-quiet scores (presentation level 65 dB SPL; phoneme scores) as a function of the hearing loss of 14 VSB-Amade/Samba users (squares). The drawn line is the best fit from a previous study (Verhaegen et al., 2008).

One problem of using a MEI with an open ear canal, such as the VSB, can be the interference between the sounds processed by the MEI and the direct sounds. It should be noted that due to the digitally processing of audio processors, the amplified sounds are delayed, usually by several milliseconds. This implies that the direct and processed sounds are out of phase for most frequencies, which can lead to a distorted sound. This phenomenon also occurs when fitting digital BTE devices with open earmolds (Groth & Sondergaard; 2004 and Stone et al., 2008). These studies reported that the (annoying) interference depends on (and can be adjusted by) the gain level of the device and compression settings. Studies investigating this interference with the use of MEIs have not yet been identified.

Chapter 9. Borderline cases: bone-conduction device versus middle ear implant versus cochlear implant

9.0 Introduction

During a roundtable discussion on borderline cases (CI2024, in Las Palmas), an overview was presented of Nijmegen studies that focussed on border areas between conventional hearing aids (behind-the ear device or BTE) and implantable hearing devices: cochlear implant (CI), middle ear implant (MEI) and implantable bone-conduction device (BCD).

Four borderline areas are discussed

–Sensorineural hearing loss:

  • Severe hearing loss: when to switch from a BTE to a CI?
  • If the hearing loss is moderate to severe, what is the better option: a BTE or MEI?

–Conductive or mixed hearing loss,

  • What is the best amplification option? BTE or BCD or MEI?
  • In case of mixed hearing loss with a severe sensorineural component, when should a powerful BCD/MEI be replaced by a CI?

There is an overlap between this chapter and previous chapters of this website. The structure of this chapter is such that it can be read without studying the other chapters. Unlike the others, this chapter has its own appendix with specific details.

The focus is on device’s advantage in terms of speech recognition in quiet. The main speech outcome used in this overview (and in all Nijmegen studies) is the (unilateral) phoneme recognition score in quiet obtained with monosyllables (MPS) while the monosyllables are presented at a normal conversational level (65 dB SPL; Snik, 2007; Verhaegen et al., 2008).

9.1 Sensorineural hearing loss

9.1.1 When switching from a BTE to a CI in severe/profound sensorineural hearing loss? 

The concept of ‘equivalent hearing loss’ is used (Snik et al., 1997): the speech recognition score of CI users is compared with that of a large group of BTE users with varying degrees of sensorineural hearing loss. For this purpose, data were collected of (53) BTE users with moderate to profound sensorineural hearing loss; Appendix 9.1 presents the selection criteria and the BTE-fitting procedure. Figure 9.1 shows, individual, unilateral MPS scores as a function of the PTA3 (mean hearing loss at 0.5, 1 and 2 kHz). The blue line connects the BTE users’ median MPS scores, divided into classes, according to their PTA3.  These data is used for a comparison with CI data.

Figure 9.1. The individual MPS values as a function of the PTA3 of BTE users and the median (p50) and p25 MPS data of the group of CI users (dotted lines).

MPS scores were obtained one year after cochlear implantation of 139 consecutive CI users with postlingual deafness. Inclusion criteria and details on the types of CI used are presented in Appendix 9.1. The median (or p50) and mean MPS were respectively 77% and 75%. The p25 MPS (the 25e percentile) 63%. The p50 (median) and p25 values are shown in the Figure 9.1 as dotted lines. The figure shows that the median and the p25 MPS values are ‘equivalent’ to those of BTE users with a PTA3 of 75 dB HL and 85 dB HL, respectively (expressed in multiples of 5 dB). In other words, if a CI candidate has a PTA3 > 85 dB HL, then the probability that his/her MPS with CI is higher than with his/her BTE is >75%. If his/her PTA3 is 75 dB SPL, that probability is 50%. This suggests that the more conservative approach using the p25 value (and thus 85 dB HL as borderline) may be the better criterion.

Several years later, Huinck et al. (2019) studied the benefit of relatively early implantation by comparing the MPS of two groups of CI users. Group 1 consisted of CI users who were selected according to rather conventional inclusion criteria (speech recognition with BTE less than 30% and/or PTA3 > 85 dB HL). The p50 and the p25 MPS values were 79% and 60%, so fairly similar to those used for the current analyses.

Huinck et al. used the data of Group1 for a comparison with CI users selected according to less strict criteria: MPS with BTE between 30% and 60% and/or PTA3 between 65 dB HL and 85 dB HL; their Group 2. The one-year post-implant MPS of the Group 2 was significantly higher than that of Group 1, namely 85% versus 79%. The authors suggested that relatively early implantation is beneficial, which was attributed to less ‘auditory deprivation’. This suggests that the less conservative PTA3 > 75 dB HL criterion might be the better choice.

9.1.2 Moderate to severe sensorineural hearing loss, better results with a MEI or a BTE?

Nowadays, only the Vibrant Soundbridge MEI is widely used (VSB; Med-El, Innsbruck, Austria). In Nijmegen the VSB has (only) been applied if a BTE was contra-indicated (due to chronic external otitis; (Zwartenkot et al., 2013). MPS data were gathered and used to answer the question which device is more powerful, a BTE or a VSB. The same straightforward method was performed as in the previous section; Figure 9.2 presents the MPS as a function of the PTA3 of 30 patients fitted with a VSB (for details regarding inclusion criteria and devices used, see Appendix 9.1). A best-fit line is presented; the black line. The blue line in the figure is copied from Figure 9.1, referring to BTE users. Apparently, the VSB scores are below the BTE scores. For example, at a PTA3 of 65 dB HL, the MPS with VSB is approx. 25% worse than that obtained with BTE and increases to 45% at 75 dB HL.

The figure suggests that, if the PTA3 exceeds 50-60 dB HL, a BTE should be fitted for better speech recognition, unless a BTE is contraindicated. Compared with the p50 MPS of the CI users, the ‘equivalent hearing loss’ is approx. 60 dB HL; therefore, if a BTE is contraindicated, cochlear implantation may be considered if the PTA3 exceeds 60 dB HL.

Although the capacity of the VSB is limited, it should be noted that this device is appreciated by patients with moderate sensorineural hearing loss and chronic external otitis. They can use this device all day without pain or itching (Zwartenkot et al., 2013). The Med-El Company advocates the VSB for patients with sensorineural hearing loss if a BTE cannot be used.

Figure 9.2. The individual MPS of MEI users plotted as a function of the PTA3. For reference purposes, more recent literature data are added (Maier et al., 2015). The mean MPS versus mean PTA3 is shown as a red dot. 

9.2 Conductive and mixed hearing loss

9.2.1 Which is the best option: a BTE or a ‘BCD’? 

Usually the cause of a (chronic) conductive/mixed hearing loss is aural atresia/stenosis or a chronically infected ear. In both cases, a BTE is contraindicated. However, in other aetiologies such as otosclerosis or tympanosclerosis, a BTE may be used.

For the latter aetiologies, whether a (powerful) BTE or a BCD is more effective, depends on the width of the AB-gap (gap between the air- and bone-conduction thresholds). Suppose a powerful BTE is used with a mean maximum output (MPO) of 112 dB HL and a mean gain of 50 dB. For example, if the AB-gap is 55 dB, then the effective MPO is really limited: namely 112 minus 55, so 57 dB HL, and the effective gain is -5 dB. Even the less powerful conventional BCDs can provide such a MPO and gain, suggesting that a choice can be made. Unlike BTEs, the MPO and gain provided by a BCD is not affected by the width of the AB-gap. The mean MPO of more powerful BCDs like the percutaneous BCD (pBCD) is significantly higher than that 57 dB HL (see section 9.2.2).

Mylanus et al. (1998) and de Wolf et al (2010) compared outcomes (MPS) obtained with powerful BTEs and pBCDs in (Nijmegen) patients with conductive and mixed hearing loss. All of their patients used a pBCD, which resulted in a dry ear. Therefore, a BTE could be fitted safely. They found that if the AB-gap exceeds 30-35 dB, the pBCD provided higher gain and thus better speech scores than the BTE. Below that AB-gap, the BTE provided the best result. In other words, if a BTE is not contraindicated while the mean AB-gap exceeds 35 dB, then a powerful BCD is the better option.

9.2.2 If a BCD is indicated, which type of BCD should be used? In other words, can we rank today’s BCDs?

Before discussing the capacity of BCDs, an overview of the different types available.

  1. The non-surgical, conventional BCD, which is held in place by an elastic softband, steal headband or adhesive coupling such as the Baha Softband (Cochlear BAS, Goteborg, Sweden), Ponto Softband (Oticon Medical, Askim, Sweden) and the Adhear (Med-EL, Innsbruck, Austria). Alternatively, a magnetic coupling may be used that requires a surgically placed subcutaneous magnet for coupling purposes; the Sophono device (Medtronic, Minneapolis, USA) and the ‘Baha Attract’ (Cochlear BAS; see Figure 9.3). All of these devices are called passive tBCDs (passive transcutaneous BCDs).
  2. Another transcutaneous approach consists of an actuator (the vibrator) implanted subcutaneously in the mastoid, driven by an externally worn audio processor. This type of BCD is called an active tBCD. Available devices: the ‘Bonebridge’ (Med-El; see Figure 9.3), the Osia (Cochlear BAS) and the Sentio (Oticon Medical).
  3. The percutaneous BCD (pBCD) with a skin penetrating implant, connecting the BCD to the skull without attenuation caused by the skin and subcutaneous tissue. Available devices: the Baha (Cochlear BAS) and the Ponto (Oticon Medical; see Figure 9.3).It should be noted that (only) for these pBCDs, more powerful audio processors are available besides the standard processors.
  4. The percutaneous ‘BCD’ (pBCD) with a skin penetrating implant, connecting the ‘BCD’ to the skull without attenuation caused by the damping tissue. Available devices: the Baha (Cochlear BAS) and the Ponto (Oticon Medical; see Figure 3). It should be noted that (only) for these ‘pBCDs’ more powerful audio-processors are available besides the standard processors

Figure 9.3. Different types of BCDs

These different types of BCDs are not equivalents. Unlike BTEs, all BCDs (and MEIs) have a limited capacity. For example, the loudest sound that BCDs can produce (the MPO or maximal power output) ranges between 50 dB HL and 85 dB HL, while humans can tolerate sounds up to 105-115 dB HL.

The MPO of BCDs and the VSB have been measured. Table 9.1 (copy of Table 2.2) provides an overview. The table shows that data from different studies are quite consistent (except for the passive tBCDs; not elaborated here).

Table 9.1. The MPO measured objectively in patients, per type of device

Device MPO measured in patients* Reference
Passive tBCD 45-58 dB HL Zwartenkot et al., 2014

Van Barneveld et al., 2018

Active tBCD, Bonebridge 64-67 dB HL Mertens et al., 2014; van Barneveld

et al., 2018; Choncheh et al, 2024

pBCD with standard processor Ponto/Baha 68-71 dB HL Zwartenkot et al., 2014

Van Barneveld et al., 2018

pBCD with super-power

processor Ponto/Baha

83-85 dB HL Zwartenkot et al., 2014

Van Barneveld et al., 2018

MEI (VSB) 80-82 dB HL Zwartenkot et al., 2014

Maier et al., 2023

* average value at 0.5, 1, 2 and 4kHz; range between studies is presented

What are the consequences of a low MPO? Figure 9.4 shows the audiogram of a patient with mixed hearing loss. Bone-conduction-thresholds are shown as well as loudness-discomfortable-levels (LDLs). The patient received a passive tBCD. The measured MPO values of that device are also shown. The mean bone conduction threshold (PTA4bc) is 32 dB HL and the mean MPO of the passive tBCD is 55 dB HL. As a result, the ‘auditory window’ with this BCD, is only 23 dB (55-32 dB). Normal conversational speech ranges from 125 Hz to 8 kHz at levels between 15 and 50 dB HL. So the range of normal speech is already 35 dB wide. Evidently, it is not possible to project relevant sounds properly (not only speech at normal conversational level, but also with raised voice or someone talking at some distance or the patient’s own voice and the other sounds) in the narrow ‘auditory window’ of 23 dB.

Figure 9.4. Audiogram of a patient with mixed hearing loss. As a (passive t)BCD is applied, only the bone conduction thresholds are relevant. MPO values are also displayed 

With more powerful devices such as a standard pBCD (MPO of 70 dB HL; Table 9.1), the auditory window becomes 38 dB and if a more powerful processor is used, like the Baha 5 SP, then the window becomes 55 dB. With the VSB, this ‘window’ is 50 dB. So, from an audiological point of view, pBCD is the best option, followed by the VSB device.

Thus, the range of normal speech is approx. 35 dB; this value is used to define application ranges for the different types of BCD and the VSB. We use the following assumption: for a given patient, a specific BCD can only be used if the ‘auditory window’ with the device is at least 35 dB. This implies that the highest acceptable average bone-conduction threshold is simply the mean MPO value minus 35 dB. Table 9.2 shows these data for the different devices.

Table 9.2. Upper limit of application, based on a dynamic hearing range of at least 35 dB per device type; the Maximum PTA4bc (mean bone conduction threshold)

Device MPO* Maximum PTA4bc
Passive tBCD 51 dB HL 15 dB HL
Active tBCD, Bonebridge 67 dB HL 30 dB HL
pBCD with standard processor Ponto/Baha 70 dB HL 35 dB HL
pBCD with super-power

processor Ponto/Baha

84 dB HL 50 dB HL
MEI (VSB) 81 dB HL 45 dB HL

* mean MPO over studies

Apparently, the different types of BCDs and the VSB are not equivalents; if the PTA4bc of a patient equipped with the standard pBCD is higher than 35 dB HL, an update to a pBCD with power- or super power processor can significantly improve the patient’s performance. Another option is to apply a VSB. For Bonebridge users with a PTA4bc exceeding 30 dB HL, a pBCD or VSB are the next options.

9.2.3 In progressive severe mixed hearing loss, when to switch to a CI?

Following the same procedures as used before, Figure 9.5 presents the MPS versus PTA3 (bone-conduction) of 26 patients with mixed hearing loss with a moderate to severe sensorineural component using the most powerful BCD at that time, the Baha Cordelle (Verhaegen et al., 2009). The PTA3 for bone conduction varied between 30 dB HL and 75 dB HL. For selection criteria etc. see Appendix 9.1. As the figure shows, when using the CI-p50 (77%), the ‘equivalent hearing loss’ is 60 dB HL.

Figure 9.5. Individual MPS values as a function of PTA3 of 26 users of the most powerful pBCD. The red dots present more recent data from 58 super-power pBCD users divided into 3 classes (Teunissen et al., 2024).

Verhaegen et al. (2009) presented the data of the first five Nijmegen patients switching from a Baha Cordelle to a CI. Two of them suffered from advanced otosclerosis and external otitis and three from chronic middle ear problems. The MPS of these five patients improved significantly from 10% (range 0 to 30%) with pBCD to 75% (range 60% to 90%) with the CI.

Regarding advanced otosclerosis, performance with a CI might be worse than in patients with normal cochlear anatomy (Rotteveel et al., 2010). Poor performance is caused by otosclerosis-related anomalies of the cochlea often leading to partial insertion of the electrode array. This might lead to a partial insertion. Because of poor bone quality, facial nerve stimulation might occur. Then, certain electrodes have to be switched off (Rotteveel et al., 2010). Therefore, the MPS scores of CI users with otosclerosis are worse than those of patients with a normal bony structure; the p50 and p25 MPS values for CI users with otosclerosis are 55% and 40% respectively (Rotteveel et al., 2010), thus more than 20% below those of CI users with anatomically normal cochleae (Figure 1). Using Figure 6 the otosclerosis-specific p50 and p25 (not shown) result in ‘equivalent hearing loss’ values of 70 dB HL and 75 dB HL, respectively.

9.2.4 Medical issues

Previous sections suggest that the pBCD is the best option from an audiological standpoint. That device has several specific medical advantages as well, but also a disadvantage. The medical benefits:

  1. Application of a pBCD requires surgery, but minimal invasive surgery, unlike all the other implantable options. Implantations are increasingly performed in outpatient clinics under local anaesthesia within 10-15 minutes.
  2. pBCD is the only implantable device that can be upgraded when hearing deteriorates without surgery
  3. The distortion of MRI images is limited in contrast to most other implantable BCDs and VSB.

As mentioned earlier, disadvantages of percutaneous implantation can include skin problems around the implant and implant loss. Over the years, the number of revisions has decreased, thanks to more than 35 years of experience. New implantation procedures have been introduced as well as optimized implants. Table 9.4, first line, presents Nijmegen data. All minor and major complications for each patient have been documented systematically, starting in 1988. Patients were seen at least five times in the first years after implantation and at least twice a year, later. In addition to planned visits, additional visits took place at the patient’s initiative. Dun et al. (2012) published data on medical complications of 903 successive adult patients, implanted from the start in 1988 up to 2008. Table 9.4 shows that one revision surgery has been performed per 30 years of follow-up (column 5). Wazen et al. (2008) published a similar study. They started later with Baha implantations, namely in 1998. That could be the reason why their result is better than that of Dun et al. More recently, prospective multicentre studies have been published using a specific novel, improved percutaneous implant and/or implant procedure (Nelissen et al., 2014; Kruyt et al., 2018; see Table 9.4). Indeed, further progress has been made, as relatively less revision surgeries were required.

It should be noted that the follow-up period of the last two studies is relatively short, 3 years. However, as has been shown by Dun et al., most complications occur in the first year after implantation.

The last columns of the table show data related to skin problems around the percutaneous implant (percentage of visits with some skin reaction). The severity of skin reaction was expressed in Holgers grades 1 to 4. Grade 2, red and moist tissue, requires local treatment; grades 3 and 4, more severe conditions that may require systematic antibiotics or revision surgery (Wazen et al., 2011). Indeed, skin problems remain common; however, the percentage of serious skin problems that require some type of intervention is decreasing (Holgers grades 2 to 4; last column).

Table 9.4.  Number of revisions related to the total follow-up period and the percentage of soft tissue reactions observed during regular visits

Author, year n Revisions2 Summed follow-up relatively All skin reactions3 Holgers grade > 2
Dun, 2012 9031 1452 4.267 yrs 1 in 30 yrs 15% 4,3%
Wazen, 2008 218 23 990 yrs 1 in 43 yrs 4
Nelissen, 2014 52 2 156 yrs 1 in 78 yrs 23,1% 1,8%
Kruyt, 2018 39 2 117 yrs 1 in 58 yrs 15,5% 1,5%

1only the adults; 2revision surgeries because of skin problems plus lost implants; 3Holgers grades 1 to 4; 4no Holgers grades reported.

Similar systematic prospective longitudinal studies on complications with other types of BCDs and MEIs are lacking.

9.3. Concluding remarks

Fortunately, several amplification options are available today for deaf and hearing-impaired patients for whom conventional amplification (BTE) is no longer effective or contra-indicated (CI, BCD, MEI), thanks to the investigators and the companies that market these devices!

As for the future, a major concern is the costs of these interventions; every update of an implantable device leads to a higher price. The rising cost of medical healthcare is a major problem in many countries. As early as 1998, Daniels & Sabin concluded: ’no country in the world can afford all of the medical care that providers can render to populations’ (Daniels and Sabin, 1998). In other words, there is a great need for affordable hearing devices including implants; see also the ‘World report on hearing’ (WHO, 2021).

Appendix 9.1 

Section 9.1.1. Regarding the 53 BTE users: the inclusion criteria were: age between 18 and 70 years, with flat or mildly sloping audiogram (difference in threshold between 0.5 kHz and 4 kHz less than 40 dB HL) with hearing thresholds < 110 dB HL up to 4000 Hz, using BTEs for at least 3 years. They were all fitted with state-of-the-art digital hearing devices using the NAL-NL fitting rule in the Nijmegen clinic, using one and the same validation protocol (Snik, 2007; Verhaegen et al., 2008).  

Selection criteria for the 139 CI users: postlingual deafness with normal cochlear anatomy. Age: between 18 and 70 years. Devices used were the Cochlear CI (N22, N24, Freedom) and Advanced Bionics CI (C1, 90K), almost equally divided. The patients were implanted between 1990 and 2009. Same evaluation protocol was used as for the BTE users. The data used are relatively old, however, comparable to that for a group of 52 CI users implanted in Nijmegen between 2010 and 2016, using similar conservative inclusion criteria. Their mean MPS was 79% (Huinck et al., 2019), close to MPS of reference CI group (75%).

Section 9.1.2. Regarding the 30 VSB users: inclusion criteria were: age 18-70 years, with flat or mildly sloping audiogram (difference in threshold between 0,5 kHz and 4 kHz less than 40 dB HL) and with hearing threshold < 80 dB HL up to 4 kHz, using their VSB for at least 1 year. The patients were fitted between 1998 and 2008 (unilaterally) with the AP303 or AP404 speech processor. For the device-fitting, the NAL rule was applied. The evaluation protocol was the same as used to test the BTE users (Snik, 2007; Verhaegen et al., 2008). 

Evidently, the included data are relatively old data. Maier et al. (2015) presented data of a more recently implanted group (their group 1, n=34). They were fitted with the Amade speech processor, using the NAL fitting rule. Individual data were not presented: in Figure 2, the average MPS is presented at the average pre-intervention PTA3 (red symbol). This average MPS is in agreement with the Nijmegen data (data-point on the best fit line) suggesting that the MPS values used might still be relevant. Note that Maier et al. reported the mean WRS using monosyllables; that score was expressed as a MPS, using a look-up table. 

 

Section 9.2.3. The study group comprised 26 patients who were using the most powerful ’BCD’ at that time, the Baha Cordelle (Verhaegen et al., 2008). The inclusion criteria were: age 18-70 years, with flat or mildly sloping bone-conduction audiogram and measurable bone conduction thresholds up to 2 kHz. They were all using a pBCD for more than a year. The mean MPO of the Baha Cordelle (0.5 to 4 kHz) is 82 dB HL, 1 to 4 dB below that of more recent powerful pBCDs, as presented in Table 2. 

 

Section 9.2.4. Deterioration of hearing loss in aging patients with chronic middle ear problems. 

For subjects with pure sensorineural hearing loss, ISO 7029 describes the deterioration of hearing sensitivity owing to aging, however, only for otologically normal subjects (ISO, 2017). In case of chronic middle ear problems, progression in cochlear hearing loss might follow the ISO 7029 norm, however, with an additional cochlear loss owing to either the infections or medical interventions in the past. We assessed that additional cochlear loss by comparing the bone-conduction thresholds of patients between 50 and 70 years, using a ‘BCD’, with the age-specific ISO 7029 thresholds. It is likely that the cochlear damage caused by the middle ear problems occurred in the past when the patients were still using a BTE, thus before using a ’BCD’ (e.g., Macnamara et al., 1996). 

Three studies with relevant data were identified. Table A9.1 shows that the ‘additional sensorineural hearing loss’ is approx. 20 dB (note: the ISO thresholds used were the average for those of men and women).

 

Table A9.1. Bone conduction thresholds compared to age-specific ISO 7029 norms of patients with mixed hearing loss using ‘BCDs’; the additional cochlear loss owing to chronic ear infections.

Study; first

author

n (ears) 0.5 kHz

dB

1 kHz 

dB

2 kHz

dB

4 kHz

dB

mean 

dB

Bosman, 2013 18 15  20 25 20  20 
Flynn, 2012 20 15 20  20  10  17 
Pfiffner, 2011*  44 19

*their group C; no frequency specific data available