Chapter 4. Basic considerations; fitting issues and device choice for the elderly

4.1 Introduction

As concluded in the previous chapter, the dynamic range of hearing with BCDs and MEIs is limited due to the low MPO of these devices. Of course, this affects proper fitting of these devices. Standard procedures cannot be used. An attempt is made to develop a fitting model based on an equitable acceptable partial use of the patient’s dynamic range of hearing. The audibility of normal conversational speech is an important factor in this model. The speech level of running speech at a level of 60 dB SPL varies between 20 and 50 dB HL for frequencies from 500 Hz to 6 kHz and between 15 and 45 dB HL at 250 Hz and 8 kHz (Killion and Mueller, 2010). Based on this, a new goal has been set: at least 35 dB of the patient’s ‘dynamic range of hearing’ must be available with the device (Zwartenkot et al., 2014 and Rheinfeldt et al., 2015). The idea is that 35 dB HL is enough to adequately understand conversational speech while the hearing device makes use of wide-dynamic range compression as a kind of automatic volume control. Wide-dynamic range compression increases the gain for low-level sounds while keeping the high-level sounds at loud but comfortable levels (Reinfeldt et al., 2015).

This new criterion is referred to as the ‘DR > 35 dB rule’.

Table 4.1 shows the average MPO data calculated from the data in Table 2.1. The second column indicates up to what mean bone-conduction threshold (average at 0.5, 1, 2, 4 kHz; the PTA4bc or SNHLc) the different devices can be used to comply the ‘DR > 35 dB rule’. An obvious conclusion is that passive tBCDs can only be used effectively in patients with predominant conductive hearing loss.

Table 4.1 The maximum permissible mean bone-conduction threshold for the different types of devices, assuming that the aided dynamic range of hearing is at least 35 dB

Device MPO* Maximum PTA4bc**
Passive tBCD 51 dB HL 15 dB HL
Active tBCD, Osia 80 dB HL 45 dB HL
Active tBCD, Bonebridge 67 dB HL 30 dB HL
Active tBCD, Sentio mini 70 dB HL 35 dB HL
pBCD with standard processor Ponto/Baha 70 dB HL 35 dB HL
pBCD with super-power

processor Ponto/Baha

84 dB HL 50 dB HL
MEI (VSB) 81 dB HL 45 dB HL

* taken from Table 2.2; mean values; **expressed in multiples of 5 dB

A ‘working range’ of 35 dB could be just enough for proper amplification of speech, in combination with wide-dynamic range compression. Wide dynamic range compression’ is the default choice for patients with pure sensorineural hearing loss using BTEs, who have a limited ‘dynamic range of hearing’ due to physiological conditions. When applying BCDs or MEIs, ‘wide-dynamic range compression’ is also used, although the limited ‘dynamic range of hearing’ is not caused by physiological conditions, but technical restrictions of the devices used. Obviously, the starting point when choosing an amplification option should be to choose a device with a high MPO, especially in children, see also Chapter 5.

Important, it should be noted that the maximum PTA4bc values per device type, as shown in the Table 4.1, correspond to the maximum PTA4bc (or SNHLc) values as obtained from clinical studies, shown in Figure 3.3B. Using the trend-curves indicated in that figure, the maximum SNHLc per device type was assessed, viz., about 45 dB HL for VSB, 15-20 dB HL for the two passive tBCDs and around 50 dB HL for pBCDs with super power processor. This match between Figure 3.3B and Table 4.1 validates the MPO-based ‘maximum PTA4bc-values’ for the applications of the different types of devices, as shown in the Table 4.1.

In short, today’s passive tBCDs can just be used effectively in patients with (sub) normal cochlear function, the active tBCDs in patients with PTA4bc between 30 dB HL and 45 dB HL, depending on the device used, and pBCDs in patients with PTA4bc up to 50 dB HL. Notably, most of these values are less optimistic than those claimed by the manufacturers.

4.2. Implanting elderly patients; the best solutions

Due to aging or a progressive (hereditary) cochlear hearing loss, the SNHLc deteriorates over time. The use of data on the age-related hearing deterioration makes it possible to estimate ‘years of effective use’ of a particular device. However, little is known (reported) on age-related hearing deterioration in patients with mixed hearing loss. One paper was identified; Figure 4.1 shows age-related decline in hearing of patients with mixed hearing loss, studied by Iliadou et al. (2006). The authors reported both air- and bone-conduction thresholds for patients suffering form OTSC7. The figure shows the PTA4bc as a function of age; see the first two rows. Data up to 60 years were presented; the indicated PTAbc values at 70 and 80 years are obtained by (non-linear) extrapolations. The thick red lines indicate whether a particular device can (still) be used effectively. For example, the figure shows that the Baha, with its ‘max PTA4bc’ of 50 dB HL (see Table 4.1), can be used effectively for up to about 80 years. The VSB can be successfully applied up to about 75 years.

Probably, the mildest scenario is that of a patient with a stable conductive hearing loss (assume 20 dB HL) and, additionally, presbycusis. Using the median age-related hearing decline for men (ISO 7029; 2017), Figure 4.2 is obtained. In this case, the effective use decreases to 50 years for the ‘Bonebridge’ and 75 years for VSB and the pBCDs while for patients aged 50+, the passive tBCD with magnetic coupling is an effective option up to age 60 years.

Figure 4.1. Effective device use for a patient with progressive hearing loss owing to otosclerosis (OTSC7). SNHLc stands for the PTA4bc. Sophono is a passive tBCD with magnetic coupling

Figure 4.2. Effective device use for a patient with stable conductive hearing loss component and a sensorineural component that deteriorates due to presbycusis

Long-term effectiveness is an often ignored but important factor in counselling, specifically in older patients, to avoid future surgeries where an active tBCD is replaced with a more powerful type or a MEI.

4.3 A simplified prescription procedure, taking the limited MPO values into account

This section deals unilateral fittings of BCDs and MEIs. Choosing the best device for a patient is challenging. First, a note should be made about bone-conduction thresholds. Note that when considering the application of a MEI, well-masked bone-conduction thresholds are essential to ensure that the cochlea of the ear to be implanted is sensitive enough for a successful application. On the other hand, when applying BCDs, unmasked bone-conduction thresholds of the ear to be treated should be considered. Due to the limited transcranial attenuation of the bone-conducted vibrations, the BCD will stimulate the cochlea with the best sensitivity, which can be either the ipsilateral or contralateral cochlea (Stenfelt, 2012). Therefore, the fitting for a BCD should be based on the unmasked bone-conduction thresholds.

To develop a practice-based prescription procedure, we used published data. The selected studies, used in Chapter 3, were reused. For each of the included studies, the gain was calculated as a function of frequency, referred to as the effective gain or EG. The EG is the difference between the unaided threshold (i.e. the bone-conduction threshold) and the aided threshold. Next, the EG was related to the unaided (bone-conduction) threshold. According to the well-validated NAL fitting rule (for sensorineural hearing loss), the gain ratio (in this case the EG divided by the cochlear (bone-conduction) threshold) should equal approx. 0.45 (with some corrections for the low frequencies; Dillon, 2012).

To see if that ratio has been reached, twenty of the previously selected papers could be included, which provided all the data needed for the calculations. Figure 4.3 shows the ‘gain ratios’, presented by device type and by degree of the cochlear loss (PTA4bc here called SNHLc). To address the discontinuities shown in Figures 3.2 and 3.3, the selected studies were divided into three subgroups, according to the mean SNHLc of the included patients. Group 1 consisted of studies in patients with a mean SNHLc below 25 dB HL, group 2 those with mean SNHLc between 25 and 40 dB HL and group 3 those with a mean SNHLc above 40 dB HL; see Table 4.2.

Table 4.2. Subgroups to address the discontinuities as observed in Figure 3.2

Group 1 Group 2 Group 3
PTA4bc < 25 dB HL 25 to 40 dB HL > 40 dB HL
Figure 3.2 shows: Negative EG Intermediate Positive EG
Hypothetical target for amplification Predominant conductive loss: compensate the AB gap! Intermediate ‘Compensate’ the SNHLc as if it was a pure sensorineural loss

 

1

2

3

Figure 4.3. The ‘gain ratio’ as a function of the frequency. Figure 4.3A shows the data from group 1 (see Table 4.2, note that BAHS stands for Baha and Ponto devices together), Figure 4.3B for group 2 and Figure 4.3C for group 3.

Figure 4.3 clearly shows that all the ‘gain ratios’ are below 0.45, the desired value according to the NAL rule (Dillon, 2012). The figure also shows that the ‘gain ratio’ is the highest at 2 kHz, regardless of device type and subgroup. Moreover, differences between the three groups are significant. The differences between the used hearing devices within the groups, are less pronounced.

Such data can be used to develop practice-based prescription procedure. Dillon and colleagues, adopting the NAL rule for patients with mixed hearing loss, argued that in (adult) patients, the entire AB gap doesn’t have to be compensated (Dillon, 2012, Chapter 10.4); a remaining AB gap of 20-25 dB is acceptable. Using 20 dB for a remaining AG gap, Table 4.3 shows the desired aided thresholds (and the resulting) EG as a function of the bone-conduction thresholds.

Table 4.3. Target values as a function of bone-conduction thresholds for the frequencies 1 kHz to 4 kHz

Bone-conduction threshold (dB HL) 0 10 20 30 40 50 60
Desired aided threshold (dB HL)* 20 20 20 20 22 27 33
As a result, the ‘EG’ (dB) -20 -10 -0 10 20 23 27

*either the ‘acceptable AB gap’ of 20 dB or 0.55 times the bone-conduction threshold; the highest value is listed

This practice-based prescription procedure can be applied regardless of the type of device used. The condition is that a device with sufficiently high MPO is selected, as stated in Table 4.1

Importantly, note that for pBCDs, Hodgetts and Scollie (2017) developed a dedicated prescription rule, based on the DSL fitting procedure. See also Voss et al., (2021) and Gascon et al., (2025).